Skip to main content
Press Enter
Starting at $1,828.00USD
Configure
Precision Digital

PD8-6363 ProtEX-MAX Explosion-Proof Dual Pulse Input Flow Rate/Totalizer

Offers all functionality of the PD6363, approved explosion-proof, displays flow rate and total from two pulse output flowmeters

Starting at
$1,828.00USD
Have questions? We're here to help!
Have questions? We're here to help!
Configure

The ProtEX-MAX™ PD8-6363 offers all the functionality of the ProVu PD6363 as a fully FM, CSA, ATEX, and IECEx approved explosion-proof product. It is specifically designed to display flow rate and total from two (2) pulse output (NPN, PNP, TTL, switch contact, sine wave, etc.) flowmeters. It displays these signals on a dual-line, 6-digit Sunbright sunlight readable display. The two display lines can be used to show both flow rates simultaneously, to alternate between the two inputs showing both the flow rate and total of each, or to show the result of math functions performed on the flow rates and totals. The total overflow feature allows up to 9-digit totals and grand totals to be displayed. The PD8-6363 includes a 24 VDC power supply to drive the flowmeter and can be equipped with up to four internal relays and a 4-20 mA analog output. It can be programmed and operated without opening the housing by using the built-in SafeTouch® through-glass buttons or the RS485 serial communication port with free Modbus® protocol.

               

Quick Overview

• Input: Two pulse inputs, each separately field selectable; open collector, NPN, PNP, TTL, switch contact, sine wave (coil), square wave; Modbus PV (slave)
• Display: Dual-line 6-digit, 0.60" (15 mm) & 0.46" (12 mm)
• Case: Smooth die-cast aluminum explosion-proof
• Power: 85-265 VAC or 12-24 VDC option
• Operating Temperature: -40 to 65°C
 

• SunBright Display Standard
• Rate, Total, and Grand Total for Each Input Channel
• Addition, difference, absolute difference, average, multiplication, division, min of A or B, max of A or B, draw, weighted average, and ratio math functions including total and grand total math
• Gate function for rate display of slow pulse rates
• 4-20 mA output option converts the pulse input to an isolated 4-20 mA output
• SafeTouch® through-glass button programming
• Modbus RS-485 serial communications
• Flanges for wall or pipe mounting
• Isolated 24 VDC @ 25 mA transmitter power supply
• Onboard USB and MeterView® Pro programming software

Under RangeDisplay flashes -99999
Display AssignmentThe Upper and Lower displays may be assigned to show: One or more rate channels: Channel A (Ch-A), B (Ch-B), or C (Ch-C), Toggle between rate channels: Ch-A & Ch-B, Ch-A & Ch-C, Ch-B & Ch-C, and Ch-A, Ch-B, & Ch-C, Total or grand total: Ch-A or Ch-B, Rate and total or grand total: Ch-A, Ch-B, Relay set points, Max and/or min values: Ch-A, Ch-B, or Ch-C, Toggle between any rate channel & units, Toggle between any rate/math channel & units, Total and units: Ch-A or Ch-B, Toggle between totals: Ch-A & Ch-B; Ch-A, Ch-B, and sum of Ch-A and Ch-B, Modbus input. The lower display may also be set to show engineering units or be off, with no display.
Programming MethodsFour through-glass SafeTouch buttons, four mechanical buttons behind glass, digital inputs, PC and MeterView Pro software, or Modbus registers.
F4 Digital Input Contacts3.3 VDC on contact. Connect normally open contacts across F4 to COM.
F4 Digital Input Logic LevelsLogic High: 3 to 5 VDC; Logic Low: 0 to 1.25 VDC
Max/Min DisplayMax/min readings reached by the process are stored until reset by the user or until power to the meter is cycled
PasswordThree programmable passwords restrict modification of programmed settings and two prevent resetting the totals.
Pass 1: Allows use of function keys and digital inputs
Pass 2: Allows use of function keys, digital inputs and editing set/reset points
Pass 3: Restricts all programming, function keys, and digital inputs.
Total: Prevents resetting the total manually
Gtotal: Prevents resetting the grand total manually
Non-Volatile MemoryAll programmed settings are stored in non-volatile memory for a minimum of ten years if power is lost.
RecalibrationAll ranges are calibrated at the factory. Recalibration is recommended at least every 12 months.
Power Options85-265 VAC 50/60 Hz, 90-265 VDC, 20 W max, or optional model with 12-24 VDC ±10%, 15 W max.
FuseRequired external fuse: UL Recognized, 5 A max, slow blow; up to 6 meters may share one 5 A fuse.
Isolated Transmitter Power SupplyTerminals P+ & P-: 24 VDC ± 10%. Isolated from the input at >500 V. Jumper selectable for 24, 10, or 5 VDC supply (internal jumper J4). All models transmitter supply rated @ 25mA max.
Isolation4 kV input/output-to-power line. 500 V input-to-output or output-to-P+ supply.
Overvoltage CategoryInstallation Overvoltage Category II: Local level with smaller transient overvoltages than Installation Overvoltage Category III.
EnvironmentalT6 Class operating temperature range Ta = -40 to 60°C
T5 Class operating temperature range Ta = -40 to 65°C
Max Power DissipationMaximum power dissipation limited to 15.1 W.
ConnectionsRemovable screw terminal blocks accept 12 to 22 AWG wire, RJ45 for external relays, digital I/O, and serial communication adapters.
EnclosureExplosion-proof die cast aluminum with glass window, corrosion resistant epoxy coating, color: blue. NEMA 4X, 7, & 9, IP68. Default conduit connections: Four ¾" NPT threaded conduit openings and two ¾" NPT metal conduit plugs with 12 mm hex key fitting installed. Additional conduit opening configurations may be available; verify quantity and sizes on specific device labeling during installation.
MountingFour slotted flanges for wall mounting or NPS 1½" to 2½" or DN 40 to 65 mm pipe mounting
Overall Dimensions (W x H x D)6.42" x 7.97" x 8.47" (163 mm x 202 mm x 215 mm)
Weight16.0 lbs (7.26 kg)
USB ConnectionCompatibility: USB 2.0 Standard, Compliant
Connector Type: Micro-B receptacle
Cable: USB A Male to Micro-B Cable
Driver: Windows 98/SE, ME, 2000, Server 2003/2008, XP 32/64-Bit,
Vista 32/64-Bit, Windows 7 32/64-Bit, Windows 10 32/64-Bit
Power: USB Port
Warranty3 years parts & labor
DisplayUpper display: 0.60" (15 mm) high. Lower display: 0.46" (12 mm) high. 6 digits each (-99999 to 999999), red LEDs with lead zero blanking.
Display IntensityEight user selectable intensity levels
Display Update Rate10 per second; up to 1 per 100 seconds (and is a function of Low Gate setting); Total: 10 per second (fixed)
Over RangeDisplay flashes 999999

Accessories

Request a Calibration Quote

We’re here to make calibration easy. Request a quote today and let our team give you the care, precision, and service you can trust.