Skip to main content
Press Enter
Fluke Calibration

1560 Black Stack Thermometer Readout

Accurate, expandable, & configurable thermometer readout for SPRTs, RTDs, Thermocouples and Thermistors

Have questions? We're here to help!
Have questions? We're here to help!

The 1560 Black Stack thermometer readout can be any kind of thermometer you want it to be, and it works in three distinctive ways. It’s a reference thermometer with an accredited calibration; it’s an automated calibration system reading your reference probe and sensors you’re testing; or it’s a high-accuracy data acquisition system. And it does each of these functions better than any other thermometer currently on the market.

The Stack consists of up to eight different modules that fit together to do any type of thermometry you choose. You can buy all of them, or any combination of them, and change the Stack and its functions anytime you want. Each module stacks behind the preceding one, and when you add a module, the Stack’s software automatically reconfigures itself to include all of the new functions supplied by that module. There’s nothing to take apart. No boards need to be installed. There’s no software to load, and nothing has to be calibrated. Just stack a new module onto the back of the previous modules and you’re ready to use the Black Stack and all of its remarkable features.

Fluke Calibration’s new Model 9935 Multi Channel LogWare makes the Black Stack an even more powerful data acquisition tool. LogWare provides graphical and statistical analysis of each channel you’re measuring (up to 96 with the Black Stack). And with alarms that can be customized, delayed start times, and selectable logging intervals, LogWare turns the Black Stack into the most powerful temperature data acquisition tool on the market.

• Reads SPRTs, RTDs, thermistors, and thermocouples
• Any configuration you like up to eight modules
• High-accuracy reference thermometer (to +/- 0.0013°C)
• Automates precision data acquisition

Model 1560 Base Unit Power: 100 to 240 VAC, 50 or 60 Hz, nominal; Attachable Modules: up to 8; Display: 4.25" x 2.25" LCD graphics, LED backlight, adjustable contrast and brightness; Automatic Input Sequencing: 1 to 96 channels; Communications: RS-232; Non-volatile Memory: channel sequence, probe coefficients; Minimum Sample Time: 2 seconds.
Extended Communication Module 3560The Extended Communication Module adds additional communication interface capability to the system. This module includes a GPIB (IEEE-488) interface, Centronics printer interface, and analog output. The GPIB interface connects the 1560 to a GPIB bus. GPIB can be used to control any function of the 1560 and read measurement data. The printer interface allows the 1560 to send measurement data directly to a printer. The analog output sources a DC signal (±1.25 VDC) corresponding to the value of a measurement.

Specifications

Model 1560 Base Unit

Power: 100 to 240 VAC, 50 or 60 Hz, nominal; Attachable Modules: up to 8; Display: 4.25" x 2.25" LCD graphics, LED backlight, adjustable contrast and brightness; Automatic Input Sequencing: 1 to 96 channels; Communications: RS-232; Non-volatile Memory: channel sequence, probe coefficients; Minimum Sample Time: 2 s

Extended Communication Module 3560

The Extended Communication Module adds additional communication interface capability to the system. This module includes a GPIB (IEEE-488) interface, Centronics printer interface, and analog output. The GPIB interface connects the 1560 to a GPIB bus. GPIB can be used to control any function of the 1560 and read measurement data. The printer interface allows the 1560 to send measurement data directly to a printer. The analog output sources a DC signal (+/- 1.25 VDC) corresponding to the value of a measurement.

 

 

Resistance Modules

.

Input Channels

Resistance Range

Basic Resistance Accuracy

Resistance Resolution

Temp.
Range

Equivalent Temperature Accuracy†

Temp.
Resolution

Excitation Current

SPRT Module 2560

2

0 Ω - 400 Ω

+/- 20 ppm of reading

(0.0005 Ω at 25 Ω, 0.002 Ω at 100 Ω)

0.0001 Ω

–260°C -  962°C

+/- 0.005°C at 0°C

+/-0.007°C at 100°C

0.0001°C

  • Thumbnail of document 1560w ugeng0000 2 14916
    Download

    1560w ugeng0000 2 14916

    Download

Accessories

Request a Calibration Quote

We’re here to make calibration easy. Request a quote today and let our team give you the care, precision, and service you can trust.