The XB OEM transmitter consists of a cable pps probe and PCB without housing. Thanks to its compact size, high accuracy and various selection possibilities, the transmitter can be adapted to customer wishes and used practically everywhere.

• Accuracy: ±1.0 %rh / ±0.2 K at 23 °C
• Temperature limit at probe: max. -100…200 °C / 0…100 %rh
• Peak load: For 100 hours between 190 °C and 200 °C. Maximal permissible continuous load: 190 °C.
• Range of application electronics: -40…85 °C / 0…100 %rh
• Metal housing (aluminum diecast) and probe of PPS
• Various probe lengths available
• Use as a simulator for system validation
• Adjusted at 23 °C and 10, 35, 80 %rh
Humidity SensorROTRONIC HYGROMER® HT-1
Temperature Sensor PT100 1/3 Class B
Accuracy at 23 °C ±1.0 %rh / 0.2 K
Resolution AirChip3000 <0.02 %rh / 0.01 K
Long-term Stability Humidiy Sensor <1 %rh / year
Measurement Range -100...200 °C / 0...100 %rh (non-condensing)
Electronics Operating Range -40...85 °C / 0...100 %rh
Output Signals Yes
Output Signals (analog) Freely scalable by user 0...1 V, 0...5 V, 0...10 V, 0(4)...20 mA
Power Supply 15...40 VDC / 12...28 VAC
Circuit Type 3 wire
Load per Analog Output V-Signal: ≥1 kΩ/V
mA-signal≤500 Ω
Calculation Dew point or frost point, freely scalable
Load Compensation Yes
Firmware Upgradeable Yes, via HygroSoftsoftware
Sensor Diagnostics (drift, state) Programmable. Default: Off
Humidity Adjustment 1-point or multi-point humidity adjustment / calibration
Temperature Adjustment 1-point or multi-point humidity adjustment / calibration
External Memory Function for Logging Yes, 2000 data point memory
PC Interface UART Yes, HygroSoft-compatible
Cable Connections 1x M16 x 1.5, to terminals
Standards CE Conformity 2021/30/EU
Audit Trail, Electronic Records Conforms to GAMP
IP Protection & Fire Protection Class IP65 / Corresponds to UL94-HB
• Measures relative humidity
• Temperature and dew/frost point in industrial environments
• OEM transmitters in climate chambers
• Incubators and monitoring of industrial processes
  • Thumbnail of document XB32 A OEM Datasheet 15336
    Download

    XB32 A OEM Datasheet 15336

    Download
  • Thumbnail of document XB32 A OEM Manual 15337
    Download

    XB32 A OEM Manual 15337

    Download

Request a Service Quote

Get a customized quote for calibration, validation, repair, or training services designed to keep your instruments accurate and compliant.
Learn more

Browse By Brands

Uncover the brands that set the standard for quality and dependability in every industry we serve.
Learn more