S8000 Chilled Mirror Hygrometer

High-precision hygrometer combines straightforward operation w/ very high sensitivity to changes in moisture content

Have questions? We're here to help!
Have questions? We're here to help!
The S8000 precision hygrometer provides a direct measurement of dew point, temperature and pressure. Suitable for both industrial and laboratory applications, the instrument is compact and self-contained.
 
Dew-point measurement is based on the proven, fundamental optical chilled mirror measurement principle, giving long-term reliable humidity measurement over the entire operating range. The S8000 uses this measurement to offer a range of available measurement units including ppmV, absolute humidity, relative humidity and ppmW.

• ±0.1°Cdp (±0.18°Fdp) accuracy
• Fundamental, accurate and drift-free measurement
• Simple configuration and operation via touch screen interface
• Precision measurements to –60°C (–76°F) dew point
• FAST – guaranteed frost formation below 0°C (+32°C)
• Measure at pressures up to 2 MPa (20 barg)
• USB, Ethernet, RS485 or RS232 connectivity
• Datalogging to SD card or via digital comms

Operating PressureLow pressure version: 0 to 1 barg / 14.5 psig
High pressure version: 0 to 20 barg / 290 psig
Sample Flow Rate0.1 to l Nl/min (0.2 to 2.1 scfh)
Dew-Point Sensor Performance
Measurement TechnologyChilled Mirror
Accuracy±0.1°Cdp
Reproducibility±0.05°C
Measurement Range–60 to +40°Cdp (–76 to +104°Fdp)
Integrated Pressure Sensor (Optional)
Measurement Range0 to 25 bara (0 to 377 psia)
Measurement Accuracy0.25% Full Scale
Measurement Unitspsia, bara, KPa or MPag
Monitors
ResolutionUser-selectable to 0.001 dependant on parameter
Measurement Units°C and °F for dew point and temperature %RH, g/m3, g/kg, ppmV , ppmW (SF6), for calculated humidities
OutputsAnalog 3 channels, user selectable 4-20 mA, 0-20 mA or 0-1 V
Digital Modbus RTU over USB, and optional Modbus RTU over RS232, RS485 or Modbus TCP over Ethernet
Alarm Two volt-free changeover contacts, one process alarm, one fault alarm; 1 A @ 30 V DC
HMI5.7” LCD with touchscreen, white on blue graphics
Data Logging SD Card (512Mb supplied) and USB interface
SD Card (FAT-16) — 2Gb max. that allows 24 million logs or 560 days, logging at 2 second intervals
Environmental Conditions–20 to +40°C (–4 to +104°F)
Power Supply85 to 264 V AC, 47/63 Hz
Power Consumption100 V A
EMC - Class A Emissions Industrial Location ImmunityComplies with EN61236:1997 (+A1/A2/A3)
Mechanical Specifications
Dimensions187 x 440 x 343mm (7.36 x 17.32 x 13.5”) h x w x d
Weight11.4kg (25.1lb)
General
Process Connections6mm Swagelok® (MALE)
tube or 1/4” Swagelok® tube (MALE)
Detection SystemRRS Triple Detection
Remote PRT Probe (Optional)
Temperature Measurement4 wire Pt100, 1/10 DIN class B
Measurement Accuracy±0.1°C
Cable Length2m (6.6') (250m (820') max)
Flow Sensor
Measurement AccuracyTypical ±5% uncalibrated
Measurement Range0 to 1000 ml/min
Storage Temperature–20 to +50°C (–40 to +122°F)
Calibration3-point traceable in-house calibration as standard
UKAS accredited calibrations optional - please consult Michell

 • Standards laboratory reference instrument

• Transformer vacuum drying verification
• Clean/dry room environmental monitoring
• Metallurgical process monitoring
• Reference instrument for commercial calibration laboratory
• Compressed air dryer R&D
• SF6 switchgear moisture testing
• Hydrogen fuel cell R&D

Request a Service Quote

Get a customized quote for calibration, validation, repair, or training services designed to keep your instruments accurate and compliant.
Learn more

Browse By Brands

Uncover the brands that set the standard for quality and dependability in every industry we serve.
Learn more