Skip to main content
Press Enter
Starting at $1,648.00USD
Configure
Precision Digital

PD8-6100 ProtEX-MAX Explosion-Proof Strain Gauge, Load Cell & mV Meter

Offers all functionality of the PD6100, approved explosion-proof, dual-line display

Starting at
$1,648.00USD
Have questions? We're here to help!
Have questions? We're here to help!
Configure

The ProtEX-MAXTM PD8-6100 offers all the functionality of the ProVu PD6100 as a fully FM, CSA, ATEX, and IECEx approved explosion-proof product. It accepts strain gauge & load cell inputs and mV signals up to 300 mV (unipolar) and ± 250 mV (bipolar) for weight and force measurement applications. The PD8-6100 strain gauge meter's dual-line display and powerful dual-scale capability allows the measurement to be displayed in two different units of measure. Another use for the PD8-6100's dual-lines is to simply display the units of measure on the bottom line. SafeTouch® through-glass buttons allow for access to zero, tare, and other functions without the need to remove the enclosure cover. The PD8-6100 includes a 24 VDC power supply to drive the sensors and can be equipped with up to four internal relays and two 4-20 mA outputs.

                    

Quick Overview

• Input: 0-10, 0-20, ±10, ±20, ±100, ±200 mV; Modbus PV (slave)
• Display: Dual-line 6-digit, 0.60" (15 mm) & 0.46" (12 mm)
• Case: Smooth die-cast aluminum explosion-proof
• Power: 85-265 VAC or 12-24 VDC option
• Operating Temperature: -40 to 65°C
 

• SunBright Display Standard
• Selectable 5 or 10 VDC excitation power supply
• Capture or programmable tare feature
• Ratiometric operation
• Dual-scale feature - single input
• Rounding function 1, 2, 5, 10, 20, 50, or 100
• Auto-zero feature eliminates zero drift
• SafeTouch® through-glass button programming
• Modbus RS-485 serial communications
• Flanges for wall or pipe mounting
• Isolated 24 VDC @ 25 mA transmitter power supply
• Onboard USB and MeterView® Pro programming software

DisplayUpper display: 0.60" (15 mm) high. Lower display: 0.46" (12 mm) high. Both are 6 digits (-99999 to 999999), red LEDs.
Display IntensityEight intensity levels
Display Update Rate5/second (200 ms)
Over RangeDisplay flashes 999999
Under RangeDisplay flashes -99999
Display AssignmentThe upper and lower displays may be assigned to PV1, PV2, PCT (percent), max/min, alternate max & min, set points, units (lower display only), or Modbus input.
Programming MethodsFour through-glass SafeTouch buttons, four mechanical buttons behind glass, digital inputs, PC and MeterView Pro software, or Modbus registers.
F4 Digital Input Contacts3.3 VDC on contact. Connect normally open contacts across F4 to COM.
F4 Digital Input Logic LevelsLogic High: 3 to 5 VDC; Logic Low: 0 to 1.25 VDC
Noise filterProgrammable from 2 to 199 (0 will disable filter)
Filter BypassProgrammable from 0.1 to 99.9% of calibrated span
RecalibrationCalibrated at the factory. Recalibration is recommended at least every 12 months.
Max/Min DisplayMax / min readings reached by the process are stored until reset by the user or until power to the meter is turned off.
PasswordThree programmable passwords restrict modification of programmed settings.
Non-Volatile MemoryAll programmed settings are stored in nonvolatile memory for a minimum of ten years if power is lost.
Input Power Options85-265 VAC 50/60 Hz, 90-265 VDC, 20 W max, or optional model with 12-24 VDC ±10%, 15 W max.
FuseRequired external fuse: UL Recognized, 5 A max, slow blow; up to 6 meters may share one 5 A fuse.
Normal Mode RejectionGreater than 60 dB at 50/60 Hz
Isolation4 kV input/output-to-power line. 500 V input-to-output or output-to-P+ supply.
Overvoltage CategoryInstallation Overvoltage Category II: Local level with smaller transient overvoltages than Installation Overvoltage Category III.
EnvironmentalT6 Class operating temperature range Ta = -40 to 60°C
T5 Class operating temperature range Ta = -40 to 65°C
Max Power DissipationMaximum power dissipation limited to 15.1 W. See PD8 instruction manual for additional details.
ConnectionsRemovable screw terminal blocks accept 12 to 22 AWG wire, RJ45 for external relays, digital I/O, and serial communication adapters.
EnclosureExplosion-proof die cast aluminum with glass window, corrosion resistant epoxy coating, color: blue. NEMA 4X, 7, & 9, IP68. Default conduit connections: Four ¾" NPT threaded conduit openings and two ¾" NPT metal conduit plugs with 12 mm hex key fitting installed. Additional conduit opening configurations may be available; verify quantity and sizes on specific device labeling during installation.
MountingFour slotted flanges for wall mounting or NPS 1½" to 2½" or DN 40 to 65 mm pipe mounting
Dimensions (W x H x D)6.42" x 7.97" x 8.47" (163 mm x 202 mm x 215 mm)
Weight16.0 lbs (7.26 kg)
USB ConnectionCompatibility: USB 2.0 Standard, Compliant
Connector Type: Micro-B receptacle
Cable: USB A Male to Micro-B Cable
Driver: Windows 98/SE, ME, 2000, Server 2003/2008, XP 32/64-Bit,
Vista 32/64-Bit, Windows 7 32/64-Bit, Windows 10 32/64-Bit
Power: USB Port
Warranty3 years parts & labor
  • Thumbnail of document Pd8 6100 datasheet 8576
    Download

    Pd8 6100 datasheet 8576

    Download
  • Thumbnail of document Provu modbus register tables 8578
    Download

    Provu modbus register tables 8578

    Download
  • Thumbnail of document LIM8 6100 0 14275
    Download

    LIM8 6100 0 14275

    Download

Accessories

Request a Calibration Quote

We’re here to make calibration easy. Request a quote today and let our team give you the care, precision, and service you can trust.
Learn more

Browse By Brands

Uncover the brands that set the standard for quality and dependability in every industry we serve.
Learn more