 The OM-EL-USB-TC thermocouple data logger measures and stores over 32,000 temperature readings from either a Type J, Type K or Type T thermocouple which plugs into a miniature female thermocouple receptacle at the base of the unit. The user can easily set up the initial data logging parameters including thermocouple type, logging rate, start-time, high/low alarm settings, logging mode and desired temperature units (°C or °F) and also download the stored data by plugging the module straight into a PC’s USB port and running the easy-to-use Windows software.

 
Downloaded data can then be graphed, printed and exported to other applications such as Excel. The usb thermocouple data logger is supplied complete with a long-life lithium battery. Data logger status is indicated by flashing red, green and orange LEDs.
 
This thermocouple data logger can be utilized in a variety of applications ranging from laboratory temperature measurement to in-field testing. Its compact size make it an ideal choice for any portable application.
 
• Plugs Directly Into USB Port for Easy Setup and Data Download—No USB Cable Required 
• Accepts Type J, K or T Thermocouples 
• 2 User-Programmable Alarm Thresholds 
• Bright Red, Green and Orange LED Indication 
• Low Battery Warning
DimensionsSee PDF
Weight43 g (1.5 oz)
Battery Life6 months (at 25°C and 1 minute logging interval)
Power1/2 AA 3.6 V lithium battery (included)
SoftwareWindows® 2000/XP/VISTA/7 (32-bit and 64-bit)
Visual Indicators (LEDs)2 LEDs; the first LED flashes orange to indicate a problem
condition with the data logger such as low battery; the second LED indicates alarm status
and flashes green (temperature within limits) or red (temperature out of limits)
Operating Temperature Range-10 to 40°C (-14 to 104°F)
Start Date/TimeSelectable in software
High/Low AlarmsSelectable in software
Logging Interval1 sec, 10 sec, 1 min, 5 min, 30 min, 1 hr, 6 hr, 12 hr
(selectable in software)
Memory32,000 readings
Temperature Units°C or °F selectable in software
Thermocouple ConnectionFemale subminiature thermocouple connector
Accuracy±1.0°C (±2.0°F)
Resolution0.5°C (1°F)
Type T-200 to 350°C (-328 to 662°F)
Type K-200 to 1300°C (-328 to 2372°F)
Type J-130 to 900°C (-202 to 1652°F)
Temperature Measurement Range: