Skip to main content
Press Enter
OMEGA

OM-CP High Temperature Data Logger w/Flexible Probe

High Temperature Data Logger With Flexible Probe

Normally ships in
3-4 weeks
Have questions? We're here to help!
Have questions? We're here to help!

Supported Applications

The Omega OM-CP-HITEMP140-FP Series is a high-temperature data logger featuring a long, flexible RTD probe with a narrow diameter. The series is designed for mapping, validation, and monitoring of high-temperature surfaces and environments. Documented applications include steam sterilization and lyophilization processes.

Operating Conditions & Performance

The data logger records up to 32,767 time-stamped readings using non-volatile solid-state memory with a resolution of 0.01°C (0.02°F). The probe measurement range is -60 to 260°C (-76 to 500°F), while the operating temperature range for the device body is -40 to 140°C (-40 to 284°F).

When equipped with an optional thermal shield, the operating environment extends from -200 to 250°C (-328 to 482°F). This extended range is time-limited; exposure duration depends on ambient temperature and whether the sensor is in air or liquid. For example, at 150°C (302°F), maximum exposure times are listed as 59 minutes for vented shields in air versus 66 minutes for flush top shields in air.

Response time specifications vary by medium:

• In Air: t60 is 30 seconds; t90 is 70 seconds.
• In Water: t60 is 3.5 seconds; t90 is 6.5 seconds.

The device maintains an accuracy of ±0.1°C (±0.18°F) and features a PFA coated flexible probe with wetted materials consisting of PFA insulated cable. The enclosure rating is IP68, and the housing material for the thermal shield options includes 316 SS components.

Configuration Options

The series supports configuration across multiple dimensions including probe length, thermal shielding type, and certification levels. Model numbers follow the format OM-CP-HITEMP140-FP - <PROBE LENGTH> - <THERMAL SHIELD> - <CERTIFICATION>. Available configurations include:

• Probe Lengths: 91.4 cm (36") or 183 cm (72").
• Thermal Shields: Vented thermal shield (OM-CP-HITEMP140-FP-TSK) and flush top thermal shield (OM-CP-HITEMP140-FP-TSK-FL). The vented enclosure measures 109 x 51 mm dia. (4.3 x 2.0") with a weight of 270 g (9.5 oz), while the flush top enclosure is 70 x 51 mm dia. (2.75 x 2.0") weighing 190 g (6.7 oz).
• Certifications: Optional NIST calibration certificate and optional 5 YEAR WARRANTY.

Data retrieval requires the OM-CP-IFC400 USB docking station or the OM-CP-IFC406 multiplexer interface. The device is powered by a user-replaceable 3.6V high-temperature lithium battery (OM-CP-BAT110). Additional accessories include the OM-CP-MULTIMOUNT-Z mount/stand and the OM-CP-SVP-SYSTEM software validation package.

Key Product Differences

Variants within the series are distinguished by probe length, thermal shield configuration, and inclusion of certification documents. The base models (e.g., OM-CP-HITEMP140-FP-36) feature a flexible probe without an external thermal shield enclosure. Models with "TSK" in the part number include a vented PTFE enclosure to extend operating limits up to 250°C (482°F). The "FL" suffix denotes a flush top thermal shield configuration, which offers different exposure time tolerances compared to the vented version at elevated temperatures. Suffixes ending in "-CERT" indicate inclusion of an optional NIST calibration certificate.

• PFA Coated Flexible Probe available in 91.4 cm (36") or 183 cm (72") Lengths 

• Operating Temperature range: -40 to 140ºC (-40 to 284ºF) 

• Optional Vented or Flush Thermal Shield for Operation up to 250ºC (482ºF) 

• Probe Measures from -60 to 260ºC (-76 to 500ºF) 

Battery lifeup to 1 yr
Battery Size and Type3.6V Li-ion
CalibrationDigital calibration through software
Data InterfaceUSB
Data Interface NoteOM-CP-IFC400 USB docking station or OM-CP-IFC406 multiplexer interface required; 125,000 baud
Diameter24.6 mm
Enclosure RatingsIP68
Form factorHandheld
Height75 mm
Measurement TypeTemperature
Memory32767
Memory Note32,767 readings
Number of Measurements1
Operating Temperature, Max140 °C
Operating Temperature, Min-40 °C
Probe Diameter0.1 in
Readings in Trigger Settings Mode10,922 readings
Real Time RecordingMay be used with PC to monitor and record data in real time
Relative Humidity, Max1
Relative Humidity, Min0%
Samples,Max32700
Sampling Rate4 Hz (4 updates/second)
Sampling Rate Note4 readings per second up to 1 reading every 24 hours
Series IDOM-CP-HITEMP140-FP-Series
Software CompatibilityWindows XP SP3/Vista/7 and 8 (32- and 64-bit)
Supply Power TypeBattery
Temperature Accuracy± 0.1 °C
Temperature Resolution0.01 °C
Temperature Sensor TypeRTD
Temperature Sensor Type NoteFlexible 100 Ω platinum RTD
Time Accuracy± 1 minute/month
Weight85 g
  • Download

    OM CP HITEMP140 FP 16480

    Download

Request a Calibration Quote

We’re here to make calibration easy. Request a quote today and let our team give you the care, precision, and service you can trust.
Learn more

Browse By Brands

Uncover the brands that set the standard for quality and dependability in every industry we serve.
Learn more