OM-CP-BRIDGE101A-30MV Compact USB Bridge/Strain Gage Data Logger w/large storage
Bridge/Strain Gage Data Logger
Supported Applications
The Omega OM-CP-BRIDGE101A is a battery-powered, stand-alone data logger designed to measure and record voltage signals from strain gages, load cells, and other low-level DC sources. The series supports monitoring of stress, torque, strain, pressure, and data acquisition from various sensors and transducers.
Operating Conditions & Performance
The OM-CP-BRIDGE101A operates in ambient temperatures ranging from -40 to 80°C (-40 to 176°F) with a humidity rating of 0 to 90% RH non-condensing. The device features an IP65 enclosure rating and is constructed from ABS plastic, weighing 24 g (0.8 oz). Data logging rates are configurable from 4 Hz down to one reading every 24 hours.
Performance specifications vary by input range:
• ±30 mV Range: Resolution of 1 μV; Calibrated Accuracy ±0.01% FSR; ±3 microvolts
• ±150 mV Range: Measurement Range ±160 mV; Resolution 5 μV; Calibrated Accuracy ±0.01% FSR; ±16 microvolts
• ±1000 mV Range: Measurement Range ±1200 mV; Resolution 50 μV; Calibrated Accuracy ±0.01% FSR; ±120 microvolts
All ranges utilize a reference voltage of 2.5V and an input range of 0 to 2.5V. Time accuracy is specified as ±1 minute/month at 20°C (68°F) for stand-alone data logging.
Power & Storage
The unit is powered by a user-replaceable 3.6V lithium battery included with the device. Typical battery life is 10 months at a 1-minute rate with a 350Ωohm load, or 2 years at a 1-minute rate with a 1000 ohm load. The logger stores up to 1,000,000 readings in non-volatile solid-state memory and supports software-configurable memory wrap of 333,000 readings in multiple start/stop mode.
Configuration Options
The series offers configurable input ranges (±30 mV, ±150 mV, or ±1000 mV) and optional NIST calibration certificates. Start modes include immediate start, delay start up to 24 months, and multiple pushbutton start/stop functionality.
Enclosure & Environmental Options
An optional weatherproof enclosure is available for harsher environments:
• OM-CP-WATERBOX101A: Weatherproof NEMA 4 (IP65) enclosure constructed from anodized aluminum with a polycarbonate cover. Dimensions are 62 H x 93 W x 25 mm D (2.45 x 3.66 x 0.93"); weight is 140 g (5 oz). Operating temperature for this enclosure configuration is -20 to 80°C (-4 to 176°F).
• OM-CP-WATERBOX101A-KIT: Maintenance kit for the OM-CP-WATERBOX101A.
Data Retrieval & Software
Data retrieval is performed via a USB connection to a PC (interface cable required) at 115,200 baud. The series requires Windows software for operation:
• OM-CP-IFC200: Includes Windows software and a 1.8 m (6') USB interface cable.
The software supports XP SP3, Vista, 7, and 8 (32- and 64-bit) systems, displaying data graphically or in tabular format with export to Excel capabilities. The device features field-upgradeable firmware and optional password protection.
Key Product Differences
The series is distinguished by input range selection and certification options:
• OM-CP-BRIDGE101A-30MV / -CERT: Bridge/strain gage data logger with ±30 mV range; the CERT variant includes a NIST calibration certificate.
• OM-CP-BRIDGE101A-150MV / -CERT: Bridge/strain gage data logger with ±150 mV range; the CERT variant includes a NIST calibration certificate.
• OM-CP-BRIDGE101A-1000MV / -CERT: Bridge/strain gage data logger with ±1000 mV range; the CERT variant includes a NIST calibration certificate.
Status indicators include Green and Red LEDs to denote logging rates, delay start mode, low battery/full memory conditions, or alarm states. A replacement 3.6V lithium battery (OM-CP-BAT105) is available as an accessory.
| Approvals | CE |
| Battery life | 10 mo |
| Battery Size and Type | 3.6V Li-ion |
| Baud Rate | 115200 |
| Calibration Accuracy | Strain : ±0.01% |
| Data Interface | USB |
| Height | 64 mm |
| Humidity Range | 0 to 90% RH non-condensing |
| Input Control Signal | 0 to 2.5 V |
| LED Indicators | Green LED Blinks: 10 second rate to indicate logging; 15 second rate to indicate delay start mode;Red LED Blinks: 10 second rate to indicate low battery and/or full memory; 1 second rate to indicate an alarm condition |
| Length | 36 mm |
| Materials | ABS |
| Measurement Type | Strain |
| Memory | 1,000,000 Readings |
| Memory Note | 1,000,000 readings; software configurable memory wrap 333,000 readings in multiple start/stop mode |
| Multi Start/Stop Mode Note | To Start the Device: Press and hold the pushbutton for five seconds, the green LED will flash during this time. The device has started logging, To Stop the Device: Press and hold the pushbutton for five seconds, the red LED will flash during this time. The device has stopped logging, Real Time Recording: The device may be used with PC to monitor and record data in real-time |
| Number of Measurements | 1 |
| Operating Temperature, Max | 80 °C |
| Operating Temperature, Min | -40 °C |
| Measurement Pressure, Max | 0 psi |
| Measurement Pressure Type | N/A |
| Process Connection Size | N/A |
| Process Connection Type | N/A |
| Real Time Recording | The device may be used with PC to monitor and record data in real-time |
| Sampling Rate | 4 Hz (4 updates/second) |
| Sampling Rate Note | 4 Hz to 1 every 24 hours |
| Series ID | OM-CP-BRIDGE101A-Series |
| Software | Windows XP SP3 / Vista / 7 and 8 (32 and 64 - bit) |
| Supply Power Type | Battery |
| Time Accuracy | ±1 minute/month |
| Weight | 24 g |
| Width | 16 mm |