Skip to main content
Press Enter
Michell Instruments

LQD-MON/SYS LIQUIDEW EExd Moisture in Liquid Analyzer

Applies a robust sensor technology to provide direct, accurate, on-line measurements for moisture dissolved in process liquids

Have questions? We're here to help!
Have questions? We're here to help!

The Liquidew EExd offers complete trace moisture measurement, applying a robust sensor technology to provide direct, accurate, on-line measurements for moisture dissolved in process liquids. Purpose designed explosion-proof analyzer provides single-or dual-channel measurements, convenient display interface and easy installation for new projects and retrofits to existing plants.

• EExd flameproof certified - IECEx, ATEX, CCSAUS and TC TR Ex
• Complete hazardous area installation
• Single or dual-channel measurements
• 0.001 ppmW to saturation range capacity
• Pre-programmed and user-entered saturation concentration values
• Remote global access options

Main Unit Measurement Parameters
Channel ConfigurationSingle and dual channel
Moisture Contentppmw
Temperature°C and °F
Water Dew Point°C and °F
Sample Flow RateStatus indication: Low-Good-High (<0.1 to >0.3 l/min)
Sensor Technology
Moisture Sensor TechnologyMichell Ceramic Moisture Sensor
Calibration Range-100 to +20°Cdp
Measurement Range-120 to +30°Cdp (0.001 ppmw to saturation)
Accuracy Dew point: ±1°C between -59.9 & +20°Cdp
Moisture content: ±10% of reading
Dew point: ±2°C between -60 & -100°Cdp
Moisture content: ±20% of reading
Analysis temperature: Accuracy ±0.2°C
Resolution0.1°C between +20 and -80°Cdp
1°C between -80 and -100°Cdp
Temperature CoefficientAlgorithm compensation (-20 to +50°C)
Operating PressureMax. 50 barg (725 psig)
Sample FlowUp to 1 l/min (0.1 - 0.2 l/min recommended)
Weight22.5kg
Operating Supply Voltage90 to 260 V AC, 50/60 Hz, 180 W Main Unit
Operating Environment-20 to +60°C up to 95% RH
Indoor or Outdoor, shaded location
Certification
Hazardous Area CertificationATEX: II 2G Ex d IIB + H2 Gb T5 (-40°C to 44°C) T4 (-40°C to +60°C)
IECEx:IECEx Ex d IIB + H2 Gb T5 (-40°C to +44°C) T4 (-40°C to +60°C)
cSAus: Class I, Division 1, Group B, C, D, T4 Tamb -40°C to +60°C
*TC TR: TC TR Ex 1Ex d IIB + H2 T4, T3 Gb (-40°C to +60°C)
Pattern ApprovalGOST-R, GOST-K
HMI
Keypad/InterfaceCapacitive touch-screen through glass
DisplayVacuum fluorescent
Data LoggingA rolling maximum of 150 data logs is available
Each log records time, date, moisture and temperature values for each channel
5 minutes minimum and 60 minutes maximum logging intervals can be set by the user
Communications and Output Two non-isolated 4-20 mA per measurement channel
500 Ω maximum load
Range and parameter settable by user
Modbus RTU @ 9600 baud-rate
Alarms: two volt free contacts per channel; one process value and one instrument status
Also available via Modbus communications
Enclosure
TypeFlameproof EExd
ConstructionCast copper-free aluminium (IP66/NEMA 4)
FinishChromate primer, Polyester P9010 powder coated (black). Meets BS3900
General
Sample Connections1/8” NPT (ATEX / IECEx) or 1/4” NPT (CSA)(female)

Request a Service Quote