The standard Rotronic Humidity and Temperature probes can be used in application with temperature from -50…100°C / 0…100 % RH. With an outstanding accuracy ± 0.8% RH ±0.1°C. For calibration the probes can be replaced without almost any downtime. Together with a Rotronic HF5, PF4/5 transmitter or HP32 and HP23 this probe is the first selection for many applications. Versatile Humidity Probe HC2A-S The HC2A-S are the most sold versatile probes from Rotronic and form the basis of the product portfolio. They measure humidity and temperature and calculate the dew/frost point. Applications for the Humidity Sensors HC2A-S Both humidity sensors are used for HVAC, food industry building services equipment, paper, textile and pharmaceutical industries.

• Accuracy: ±0.8 %rh, ±0.1 K, at 23 °C ±5 K
• Range of application: -50…100 °C / 0…100 %rh
• Digital interface (UART) and scalable analog outputs, 0…1 V
• GAMP compatible (HygroSoft software)
Probe typeStandard black probe
Operating range -50…100 °C / 0…100 %rh
Accuracy w. adjustment profile "standard"±0.8 %RH / ±0.1 K, at 10...30 °C; Adjustment at 23 °C and 10, 35, 80 %RH
Accuracy w. adjustment profile "high precision" ±0.5 %RH / ±0.1 K, at 10...30 °C at 23 °C and 10, 20, 30, 40, 50, 60, 70, 80, 90 %RH
Power supply/Power consumption3.3…5 VDC
Current consumption4.5 mA at 3.3 VDC
Long-term stability<1 %rh / year
Humidity sensorHygromer HT-1
Temperature sensorPT100 1/3 DIN Class B
Probe protectionPolycarbonate plastic cage, black
Filter cartridgePolyethylene
Filter typePolyethylene, 40 μm
Response time T 63<15 sec.
Allowed wind speed20 m/s, with filter
Psychrometric calculationsDew or frost point
Reproducibility airchip<0.02 %rh / 0.01 K
Exploitation softwareROTRONIC HW4
Probe adjustment by software1 Point & multipoint %rh & °C
Firmware updateWith internet and HW4 software
User informationFrom Software HW4 (devicemanager)
Data processing via HW4With interface cables
Device protection by passwordYes
Type of output signals2 x 0…1 VDC
Analogue outputs scaleable by the userYes
Analog output signal (standard)0…1 V = 0…100 %rh; 0…1 V = -40…60 °C
Digital output signal (standard)UART
Service interfaceUART
Reverse voltage protectionMechanically in the connector
Max. cable lengthUp to 5 m with passive cable. Up to 100 m with active cable
Probe materialPEEK
Enclosure materialPolycarbonate
HousingPolycarbonate
FDA/GAMP compatibilityFDA 21 CFR Part 11 and GAMP compatible
DimensionsØ15 x 108 mm
Weight10 g

 • Pharmaceutical industries

• Incoming goods inspection
• Food Industry
• Sterilization of clean rooms, incubators and other devices

  • Thumbnail of document 12 1085 0103 Hygro Clip2 Advanced web 13533
    Download

    12 1085 0103 Hygro Clip2 Advanced web 13533

    Download
  • Thumbnail of document 59055 E HC2 A datasheet 15317
    Download

    59055 E HC2 A datasheet 15317

    Download

Request a Service Quote

Get a customized quote for calibration, validation, repair, or training services designed to keep your instruments accurate and compliant.
Learn more

Browse By Brands

Uncover the brands that set the standard for quality and dependability in every industry we serve.
Learn more