EL-21CFR-ULT Compliant Ultra Low Temperature Cryogenic Vaccine Data Logger

Designed for monitoring vaccines in cryogenic dry ice storage

This product is no longer sold. View active Product

Have questions? We're here to help!
Have questions? We're here to help!

This 21CFR Part 11 compliant ultra-low temperature data logger is designed for monitoring vaccines in cryogenic dry ice storage. The measurement range is -100 to +100°C (-148 to +212°F) using the probe supplied with the logger.  The device can record over 32,000 readings and is simple to set up using the 21CFR Part 11 software, which allows users to set their own high and low alarm thresholds, with alarm hold and delayed start features as well.  Flashing lights alert users to any temperature excursions.  Just plug the logger into your PC's USB port to download data which can then be graphed, printed and exported to other applications for detailed analysis.

Used in conjunction with Lascar's 21CFR Part 11 software, the logger enjoys all of the core features of the standard EasyLog software, but with additional 21CFR characteristics including user permissions, encrypted data which cannot be edited, full audit trail and electronic signatures for all activity.
 
The product is supplied with a ½ AA battery,  the probe (300mm probe on a 1m long cable), and a mounting bracket.
 
This product complies with BS EN 12830:2018 (Temperature recorders for the transport, storage, and distribution of temperature-sensitive goods).
 
 
 
▪ Measurement range -100 to +100°C (-148 to +212°F)
▪ T-type probe on 1m cable
▪ Stores over 32,000 readings
▪ Logging rates between 1 second and 12 hours
▪ Immediate and delayed logging start
▪ User-programmable alarm thresholds 
▪ Status indication via red and green LEDs
▪ Use as part of a 21CFR Part 11 compliant system
▪ EasyLog 21CFR software available as a free download
Dimensions (logger)118 x 27 x 27mm (4.6 x 1.1 x 1.1”)
Calibration CertificateAvailable Separately
Battery Life2 Years
Dimensions (probe)300 x Ø1.5mm x 1m cable (11.8 x Ø0.06" x 39.4" cable)
Operating temperature (logger only)-10 to +40°C (+14 to +104°F)
Logging RateUser selectable between 1 second and 12 hours
Readings32,510
Accuracy±2.5°C (±4.5°F) typical
Measurement Range-100 to +100°C (-148 to +212°F)
▪ Ultra Low Temperature Vaccine Storage
▪ Vaccine Storage
▪ Pharmaceutical Cold Chain
 

  • Thumbnail of document El 21cfr quick start guide iss2 07 26 singlepageversion 12104
    Download

    El 21cfr quick start guide iss2 07 26 singlepageversion 12104

    Download
  • Thumbnail of document El 21cfr ult iss1 01 21 us version 13554
    Download

    El 21cfr ult iss1 01 21 us version 13554

    Download

Accessories

Request a Calibration Quote

We’re here to make calibration easy. Request a quote today and let our team give you the care, precision, and service you can trust.
Learn more

Browse By Brands

Uncover the brands that set the standard for quality and dependability in every industry we serve.
Learn more