DP606A Universal 6/12 Channel 1/4 DIN Panel Meter

Universal 6/12 Channel ¼ DIN Panel Meter, Universal Input

Have questions? We're here to help!
Have questions? We're here to help!

 The DP606A and DP612A Series is our newest line of affordable multi-channel panel meters. These models offer increased accuracy, wider operating temperatures and increased flexibility over the CN606 and CN612 models that they replace. The input, power and communications are also now isolated allowing for easier deployment and more reliable operation. The rugged extruded aluminum ¼ DIN case combines with the improved isolation to provide an extremely durable product.

 
These meters can read up to 12 sensors and include user selectable RS232 or RS485 communications using Modbus®. Two user configurable alarm relays are standard on all models. The universal input allows user selection between Thermocouple (9 Types), RTD 2-Wire, RTD 3-Wire, 4 to 20 mA and 1000 mV input options. These input selections are firmware selectable through the front panel buttons or Modbus serial commands. No dip switch setting changes are required.
 
Excellent fit for OEM instrumentation, automotive and aerospace test stand, factory automation, and power generation applications.
 
Note: While the inputs are universal and can be changed via firmware (no dip switches), all 6 or 12 inputs on a given unit must be set to the same type of input for operation.
 
Software: This multi-zone panel meter is most easily configured with Omega Sync, a software tool that enables users to configure device runtime parameters, view process values, and seamlessly export data. At the moment, Omega Sync supports the CN616A and DP606A/DP612A devices and we look to expand its reach to future products. For long-term data storage, our experts recommend using Omega Enterprise Gateway (OEG) that provides a powerful data historian, alarming, visualization and OPC-UA integration.
• Universal TC, RTD and Process Inputs Firmware Selectable
• User Programmable Alarms Individually Settable per Channel
• Adjustable Scan Time: 400 mS Max Scan Rate per Channel
• 2 User Assignable Alarm Relays
Number of Zones61 or 12
1RTD 3 Wire 6 channels for all models
Display15 mm 7-seg LED
Display Time1 to 40s
Input Types
RTD2 or 3 wire
PT100 0.00385
CU10 0.00427
NI120 0.00672
Process4 to 20 mA, 0 to 1V
Resolution1º, 1 mV, 0.01 mA
AccuracySee Chart Below
A2D24-bit Sigma Delta
CommunicationsRS232 and RS485 (User Selectable)
Up to 115.2k baud
Modbus Protocol (RTU)
Input Power
AC Models120/240 Vac; 50/60 Hz
Isolation per 1 Min
AC Units2.3k Vac (3.2k Vdc)
DC Units1.5k Vdc
Power to Alarm Relay2.3k Vac (3.2k Vdc)
Power to Serial2.3k Vac (3.2k Vdc)
Input to Serial500 Vac (720 Vdc)
Alarm Relays
AC Models240 Vac max
5A max load
DC Models36 Vdc max
3A max load
De-energizes on Alarm Latching or Non-Latching High, Low or Hi/Low
Operating Temp-20 to 70ºC (-4 to 158ºF)
Storage Temp-40 to 85ºC (-40 to 185ºF)
Panel Cutout¼ DIN 92 x 92 mm (3.6 x 3.6")
Dimensions95 H x 95 W x 135 Dia mm (3.7 x 3.7 x 5.3")
Min Rear Panel Depth140 mm (5.5")
Weight725 g (26 oz)
Approvals
ECCNEAR99
Safety2014/35/EU
EMC2014/30/EU
RoHS2011/65/EU
UL File No.E209855
UL/cUL Listed
Power to Input
TCJ, K, T, E, R, S, B, C, N
Scan Time400 mS
DC Models9 to 36 Vdc
3 watts maximum