Skip to main content
Press Enter
OMEGA

DP400TP High Speed Temperature & Process Meter, NFC Enabled

High Speed Temperature and Process Meter, NFC Enabled

Normally ships in
2-3 weeks
Have questions? We're here to help!
Have questions? We're here to help!

 Supported Applications

The Omega DP400TP is an indicator and panel meter designed for the acquisition and retransmission of process variables, functioning as a fast transient recorder. The device supports temperature measurement via thermocouples (types K, S, R, J, T, E, N, B) and thermoresistance sensors (PT100, PT500, PT1000, Ni100, PTC1K, NTC10K). It also accepts linear voltage/current inputs (0-10 V, 0-20 mA, 4-20 mA, 0-60 mV) and potentiometer inputs (6 KΩ, 150 KΩ).

Key functional capabilities include:

• Alarm management via relay outputs.
• Analog retransmission of process values or setpoints.
• Data visualization with bar graph and process trend displays featuring programmable sampling time.
• Mathematical functions including Totalizer and Sum calculations.

Operating Conditions & Performance

The DP400TP operates within an ambient temperature range of 0-40 °C with a humidity rating of 35..95 uR%. The enclosure provides IP54 protection on the front panel (with gasket) and IP20 for the box and terminals. The housing is constructed from polycarbonate V0 material.

Performance specifications include:

• Tolerance at 25 °C: +/-0.2% ±1 digit (F.s.) for thermocouple, thermoresistance, and V/mA inputs.
• Cold junction accuracy: 0.1 °C/°C with automatic compensation from 0..50 °C.
• Analog output resolution: All outputs are 16-bit with +/-0.2% (F.s.) tolerance.

The device features a 2.42” monochrome (yellow) OLED graphical display with a resolution of 128x64 pixels and supports NFC connectivity for configuration via the DP-USB-CARD accessory.

Configuration Options

• Housing & Mounting: Standard format is 96x48mm, configurable for both horizontal and vertical mounting. Weight is approximately 165 g.
• Power Supply: Extended range of 24..230 Vac/Vdc ±15% at 50/60 Hz (Consumption: 8 VA).
• I/O Configuration:
    • Relay outputs: 2 relays, contacts rated for 2 A - 250 V~ resistive charge.
    • Analog output options: 1 voltage linear (0..10 Volt) or 1 current configurable as 0..20mA or 4..20mA.
    • Digital inputs: 2 D.I. programmable via software.
• Communication: RS485 interface supporting the Modbus-RTU protocol.
• User Interface: Intuitive multilingual interface with OLED display and RFID/NFC support.

Key Product Differences

The DP400TP is identified as a single model family (DP400TP) without further configurable segments in the provided documentation. The device supports multiple input types through software configuration rather than distinct hardware variants for each sensor type.

• UL Certified
• Visualization & Multilingual Interface
• High Speed & Sampling Rate Up to 242 Hz
• Alarm and Signal Re-transmission Functions
Additional Features 1Trend visualization
Control Output2 SPDT Relay
Cutout Dimensions48 x 96 mm
Data InterfaceRS485
Data Interface NoteRS485 Modbus RTU - Slave (1200 ~ 115200 Baud) galvanically isolated from Power supply/Inputs/Outputs
Display TypeGraphical OLED
Electrical ConnectionScrew Terminal
Enclosure MaterialPolycarbonate V0
Enclosure RatingsIP65, NEMA Type 4
Form factor1/8 DIN
Input Type Note1 Analogue: Res.16 bit, Selecting for T/C type K, S, R, J, T, E, N, B (automatic compensation of the cold junction 0 ~ 50 °C, ± 0,2% F.S. ± 1 Digit F.S.), thermoresistances PT100, PT500, PT1000, Ni100, PTC1K, NTC10K (ß 3435K), process signals 0-10 V(54000 points), 0/4-20mA(40000 points), 0-60 mV (16000 points), Potentiometer 6 KO, 150 KO (50000 points) Customizable linear input (max 16 steps) 2 Digital: PNP inputs programmable for Run / Hold / Tare-Zero / Alarms reset/ Peaks reset/ Totalizer reset / Activate-Reset Sum / Parameters-Setpoint Lock
Number of Inputs1
Operating Temperature, Max45 °C
Operating Temperature, Min0 °C
Output Configurations2 Analog Current
Output Configurations Note1 output 0 ~ 10V (60000 points) - 1 output 0/4 ~ 20mA (60000 points) Analogue Retransmission: Process values / Setpoints
Output Detail2A - 250VAC (resistive charge)
Power Consumption8 VA
Quick Set-up OptionsSoftware LABSOFTVIEW and/or Memorycard
Relative Humidity35 ~ 95 RH% (non condensing)
Sampling Rate4.12 Hz to 242 Hz (4.12 to 242 updates/second)
Sensor/Transmitter Power24 VDC - 30mA for external sensors supply (loop-powered)
Series IDDP400TP-Series
Supply Power Note24 ~ 230VAC/DC ±10% 50/60 Hz (galvanic isolation 2500V)
Supply Power TypeAC/DC
Supply Voltage24 to 230 Vac/Vdc
Thermocouple TypeJ, K, T, E, N, R, B, S
Weight165 g

Request a Calibration Quote

We’re here to make calibration easy. Request a quote today and let our team give you the care, precision, and service you can trust.
Learn more

Browse By Brands

Uncover the brands that set the standard for quality and dependability in every industry we serve.
Learn more