Skip to main content
Press Enter
OMEGA

DBCL-400 Ambient + 5°C to 450°C Temperature Dry Block Calibrator

Portable Dry Block Calibrator with Multiple Well Insert

Normally ships in
3-4 weeks
Have questions? We're here to help!
Have questions? We're here to help!
Configure

Supported Applications

The Omega DBCL-400 Ambient 5 °C to 450 °C Temperature Dry Block Calibrator provides a safe, dry, constant temperature source for checking and calibrating a wide range of temperature sensors, systems, indicators, and thermometers. Designed as both a benchtop unit and a portable field instrument weighing 11 lbs (5 Kg), the series utilizes a machined aluminum block as the heat transfer medium to establish stable thermal environments.

Operating Conditions & Performance

The DBCL-400 operates within an ambient temperature range of 5°C/9°F to 40°C/104°F and supports a measurement range from 5°C/9°F above ambient up to 450°C/850°F. An over-temperature limit protection circuit is set at 470°C/875°F. The unit features an automatic fan cooling system with the following thermal response characteristics:

• Heat up time (25°C to 400°C): 12 minutes
• Cool down (400°C to 100°C): 20 minutes

Accuracy specifications vary by temperature range: ±0.4°C (50 to 400°C) / ±0.7°F (122 to 752°F), and ±0.6°C (400 to 450°C) / ±1.0°C (752 to 850°F). Stability after 15 minutes is specified as ±0.050°C (50 to 400°C) / ±0.090°C (122 to 752°F). Well-to-well radial uniformity is documented at 0.015°C at 100°C and 0.052°C at 300°C. The display resolution is 0.1°. Quoted specifications apply for an ambient temperature range of 10°C/50°F to 30°C/86°F; performance may deteriorate outside this range while maintaining safe operation.

Configuration Options

The series supports two primary power supply configurations: 230V, 50/60Hz, 900W and 120V, 50/60Hz, 900W. Electrical connections utilize a power cable with plug; wire color codes are Live (Brown for 240V / Black for 120V), Neutral (Blue for 240V / White for 120V), and Earth ground (Green/yellow or Green). The unit requires earth grounding to ensure proper electrical safety.

Configuration options include various inserts, accessories, and replacement parts:

• Inserts: Multiwell insert with 1/8", 3/16", 1/4", 5/16" & 3/8" holes (DBCL-400-3041); Blank insert (DBCL-400-3047); Inserts configured for specific hole combinations including 5 x 1/4" (DBCL-400-3043), mixed sizes such as 1 x 9/16" & 1 x 1/4" (DBCL-400-3048), and Blackbody Source insert for IR pyrometers (DBCL-400-3129).
• Accessories: Unit carrying strap (4168), Insert extractor (4153), Carrying case (DBCL-3052), and region-specific power cables including UK 240V with 13amp plug (DBCL-UKCABLE) and Euro style R/A Schuko plug (4164).
• Replacement Parts: Heaters, temperature controllers, PRTs, solid state relays, fuses (4 amp for 240V units; 8 amp for 120V units), heater blocks, and cooling fans.

The unit dimensions are 8.75 x 8 x 8 inches / 222.25 x 203.2 x 203.2 mm (excluding carrying strap). Immersion depth is specified as 4.5" (114.3mm).

Key Product Differences

The DBCL-400 series distinguishes between two voltage variants: the DBCL-400-120 for 120V operation and the DBCL-400-240 for 230V/240V operation. Units marked 230V on the rating plate operate at 220V, while units marked 120V operate at 110V; in both cases where voltage deviates from nominal ratings, the heating rate degrades by approximately 8%. The series includes a fused plug for UK use fitted with a 5 AMP fuse. No operator-maintainable parts are located within the equipment.

• Maximum temperature of 450°C (850°F)
• Minimum temperature of 5°C (9°F) above ambient
• An independent over-temperature cutout, 470°C (875°F)
• Heat up time 25° C to 400°C with stability at 12 minutes
Accuracy± 0.4 °C
Accuracy Detail±0.4°C (50 to 400°C) ±0.7°F (122 to 752°F) ±0.7°C (400 to 450°C) ±1.3°F (752 to 850°F)
Cool-Down Time NoteCool down 400°C to 100°C: 20 minutes
Dimensions Note8.75 x 8 x 8 inches/222.25 x 203.2 x 203.2 mm *excluding the carrying strap
Display Resolution0.1°
Frequency50/60 Hz
Heat-Up Time NoteHeat up time 25° C to 400°C: 12 minutes
Height222.25 mm
Length203.2 mm
Operating Temperature, Max40 °C
Operating Temperature, Min5 °C
Power Consumption900 W
Process Temperature, Max450 °C
Process Temperature, Min5 °C
Process Temperature Note5°C/9°F above ambient to 450°C/850°F;Temperature Uniformity: 0.020°C at 200°C & 0.030°C at 400°C
Relative Humidity, Max0.95
Relative Humidity NoteUp to 95% relative humidity, non-condensing
Series IDDBCL-400
Temperature Overload ProtectionFan Cooling: Automatic; Overtemperature Limit: 470 °C
Temperature Stability±0.050°C (50 to 400°C) ±0.090°F (122 to 752°F) (After 15 minutes)
Weight5 kg
Width203.2 mm
  • Download

    DBCL 400 Manuals 16217

    Download

Request a Calibration Quote

We’re here to make calibration easy. Request a quote today and let our team give you the care, precision, and service you can trust.
Learn more

Browse By Brands

Uncover the brands that set the standard for quality and dependability in every industry we serve.
Learn more