Skip to main content
Press Enter
OMEGA

CT5100 1 or 2 Channel Programmable Process Input Circular Recorders

1 or 2 Channel, 10" Circular Chart Recorders

Normally ships in
2-3 weeks
Have questions? We're here to help!
Have questions? We're here to help!
Configure

Supported Applications

The CT5100 Series are programmable 1- or 2-channel circular chart recorders that measure and display up to two process variables and trace them, one or two pens, onto a 254 mm (10") circular chart. A 4-digit LED display and three front-panel keys make the recorder easy to configure and read at the point of use.

A single recorder accepts a wide mix of process inputs - thermocouple, RTD, DC current and DC voltage - so it suits temperature, flow, level and general process trending in the field or on a panel. Up to two alarm points per pen can drive SPDT relay contacts for on-off control, and an RS-485 port running MODBUS RTU protocol allows the recorder to be integrated into a wider monitoring system.

Housed in an injection-molded Noryl case with an acrylic window and a NEMA 3 (IP54) rating, the recorder is UL and cUL listed and FM Approved. Note: installation should be performed by qualified personnel, away from excessive moisture, oil, dust and vibration, and the required AC input power should be verified before installation.

Operating Conditions & Performance

The recorder accepts thermocouple types J, K, T, R and S and Pt100 RTDs, with reference ranges of:

• Type J: 0 to 760 °C (0 to 1400 °F)
• Type K: 0 to 1360 °C (0 to 2500 °F)
• Type T: -200 to 400 °C (-330 to 752 °F)
• Types R and S: 200 to 1650 °C (400 to 3000 °F)
• Pt100 RTD: -140 to 400 °C (-220 to 752 °F)

Process inputs cover DC current of 0 to 20 mA and 4 to 20 mA, and DC voltage spans of 0 to 25 mV, 0 to 50 mV, 10 to 50 mV, 0 to 5 V and 1 to 5 V. Accuracy is ±0.25% of span ±1 degree for thermocouple and RTD inputs and ±0.25% of scaled span plus 1 least-significant digit for mA, mV and Vdc inputs, with a cold-junction compensation error of ±0.2°C @ 25°C (±0.36°F @ 77°F). Input impedance is under 0.1 Ω for thermocouple and Pt100 inputs.

The recorder scans once per second for non-RTD inputs and once per 1.2 seconds for RTD inputs. It operates from 0 to 50°C (32 to 122°F) at 10 to 90% RH non-condensing and stores at -40 to 65°C (-4 to 149°F). Standard power is 90/264 Vac, 50/60 Hz, with a low-voltage option of 20 to 50 Vac at 50/60 Hz or 22 to 65 Vdc. Alarm relay contacts are rated 5 A resistive @ 115 Vac and 2.5 A resistive @ 230 Vac.

Configuration Options

Models differ by pen count and by the alarm, security and power options fitted. The core distinction is channel count:

• 1-pen recorders (CT5100 family) trace a single process variable.
• 2-pen recorders (CT5110 family) trace two variables at once, with alarm relays configurable per channel.

Within each channel count, alarm capability steps up from none, to one alarm relay, to two alarm relays per channel, and an alarm can be configured as a latching limit device that requires a manual reset button. Add-on options are ordered by suffix: +CE adds CE approval and +XPS adds a transmitter power supply, while a door-lock version secures the front door. A low-voltage power option covers 20 to 50 Vac or 22 to 65 Vdc where mains power differs.

Key Product Differences

The model number follows the guide CT5100 - <recorder type> - <alarm relays> - <door lock>. The orderable line covers both channel counts and their alarm and lock builds:

• CT5100 - 1-pen recorder.
• CT5101 - 1-pen recorder with 1 alarm relay.
• CT5102 - 1-pen recorder with 2 alarm relays.
• CT5100-LOCK - 1-pen recorder with door lock.
• CT5110 - 2-pen recorder.
• CT5112 - 2-pen recorder with 1 alarm relay per channel.
• CT5114 - 2-pen recorder with 2 alarm relays per channel.
• CT5110-LOCK - 2-pen recorder with door lock.

Consumables and accessories include red pens for channel 1 (CT7000-RED), green pens for channel 2 (CT7000-GREEN), and circular chart paper in 24-hour and 7-day formats. Each recorder measures 355.6 W x 355.6 H x 96.5 mm D (14 x 14 x 3.8"), needs a panel depth of 64 mm (2.5") with a 33 mm (1.3") protrusion, and weighs 8.8 kg (15 lb) max.

• 254 mm (10') Chart
• 1- or 2-Pen Versions,
• Thermocouple, RTD, dc Current, or Voltage Input
• Programmable Inputs
Accuracy± 0.5 %
Accuracy DetailChart Recording: 0.5% of chart span reference accuracy, Chart Rotation: ±0.5% of rotation time, assuming all backlash removed, Chart Rotation: ±0.5%, Chart Speed: ±1%
Approvals NoteSafety: UL Approved for USA:UL 1092, UL 916 and QUXY, pending UL Certified for Canada:CSA Spec 142, pending Limit/Device: FM, pending
Chart Diameter254 mm
Chart NoteChart Span: Bottom and top of span, -9999 to 9999 units; Chart Drive: DC Stepper Motor; Chart Drive: Stepper Motor
CMRR>120 dB @ 50/60 Hz, 260 Vac max
Cold Junction Compensation0.04°/°C deviation from 25°C
Conduit Entrance3 Openings on the Right Side
Current Measurement Range0 to 20 mA
Cutout Dimensions322.58 x 322.58 mm
Display Note4-digit, 14 mm H (0.56"), red, 7-segment, LED display, Display Modes: Normal: process value(s) or blank Keypad: 3 keys for programming and unit operation
Display TypeLED
Enclosure RatingsIP54, NEMA Type 3
Height355.6 mm
Impedance Input>100 MΩ for T/C and mV inputs, 100 kΩ for 5 V inputs, 4.7 Ω for µA inputs
Input NoteInput Correction: Offset adjustment: 999 to 999 units; Input Isolation: 350 Vac, 500 Vdc
LED Indicators5 Red LED Alarm Status Indicators, 1 green LED pen 2 indicator
Length96.5 mm
MaterialsNoryl, Acrylic
Measurement Range NoteThermocouple: J: 0 to 760°C, K: 0 to 1360°C, T: -200 to 400°C, R: 200 to 1650°C, S: 200 to 1650°C RTD: -140 to 400°C Current: 0 to 20 mA, 4 to 20 mA Voltage: 0 to 25 mV, 0 to 50 mV, 10 to 50 mV, 0 to 5 V, 1 to 5 V
Measurement TypeThermocouple, RTD, Current, Voltage
MountingPanel Mount or Wall Mounting
NMRR85 db @ 60 Hz or greater
Operating EnvironmentReference Operating Condition: 25 °C
Operating Temperature, Max50 °C
Operating Temperature, Min0 °C
Panel Depth64 mm
Panel Protrusion33 mm
Pen TypeDisposable Fiber tip
Power Consumption18 VA
Relative Humidity10 to 90% RH, non-condensing
RTD Type100 Ω Pt sensor; α= 0.00385
Sampling Rate1 Scan per second for Non-RTD inputs; 1 Scan per 1.2 seconds for RTD inputs
Sensor Input Configuration150 µA
Series IDCT5100-Series
Standard Limits of ErrorError Limit: Cold-Junction Compensation Error: ±0.2°C @ 25°C, Linearization Error: T/Cs: ±0.25°C typical, ±0.5°C worst case , RTDs: ±0.1°C typical, ±0.3°C worst case, Ambient Temperature Error: ±0.01% of span per °C deviation from 25°C; Measurement Error: Type J, K, T, R, S and RTD: ±0.25% of span ±1 degree; mA, mV and Vdc: ±0.25% of scaled span plus 1 least significant digit
Storage Temperature, Max65 °C
Storage Temperature, Min-40 °C
Supply Power NoteLine Voltage: 90/264 Vac, 50/60 Hz; optional: 20 to 50 Vac, 50/60 Hz, or 22 to 65 Vdc
Supply Power TypeAC
Supply Voltage90 to 264 Vac
Switch Rating5 A @ 115 Vac
Switch Rating NoteRelays: SPDT; contacts rated 5 A resistive @ 115 Vac,2.5 A resistive @ 230 Vac, 1⁄8 hp @ 230 Vac (single phase),250 VA @ 115/230 Vac
Switch TypeSPDT NO/NC, Selectable
Temperature Measurement RangeJ: 0 to 760°C, K: 0 to 1360°C, T: -200 to 400°C, R: 200 to 1650°C, S: 200 to 1650°C, RTD: -140 to 400°C
Thermocouple TypeJ, K,T, R, S
Voltage Measurement Range0 to 25 mV, 0 to 50 mV, 10 to 50 mV, 0 to 5 V, 1 to 5 V
Weight8.8 kg
Width355.6 mm
Wire Code Option2- or 3- Wire

Request a Calibration Quote

We’re here to make calibration easy. Request a quote today and let our team give you the care, precision, and service you can trust.
Learn more

Browse By Brands

Uncover the brands that set the standard for quality and dependability in every industry we serve.
Learn more