CL510A PT100 & PT1000 RTD Simulators with multiple or single curves

RTD Simulators

Normally ships in
3-4 weeks
Have questions? We're here to help!
Have questions? We're here to help!
Configure

Supported Applications

The Omega CL510A Series consists of portable RTD (Resistance Temperature Detector) simulators designed for checking and calibrating RTD instruments. These handheld devices are compatible with smart transmitters, PLCs, DCS systems, controllers, and multichannel recorders.

Operating Conditions & Performance

The series operates within an ambient temperature range of -25 to 60 °C (-10 to 140 °F). Relative humidity limits are defined as 10 % ≤ RH ≤ 90 % (non-condensing) for temperatures between 0 and 35 °C, and 10 % ≤ RH ≤ 70 % (non-condensing) for temperatures between 35 and 60 °C. The unit provides over-voltage protection rated to 60 V dc for up to 30 seconds.

Accuracy specifications vary by excitation current conditions:

• General Accuracy: ±(0.015% of Setting in Ohms + 0.05 Ohms)
• External Excitation Current (1 to 10.2 mA): ±(0.015% of Setting + 0.05 Ohms)
• Below 1 mA External Excitation Current: ±(0.015% of Setting + 0.01 mV/mA Excitation Current + 0.05 Ohms)

The device supports resistance ranges from 0.00 to 410.00 Ω and 410.1 to 4001.0 Ω. Temperature drift is specified as ± 0.05 Ohms/°C.

Configuration Options

The CL510A Series offers configurable simulation capabilities based on the selected model variant, supporting single or multiple RTD curves with specific alpha coefficients and temperature ranges:

• Sensor Types Supported: Pt 100 Ohm (DIN/IEC/JIS 1989, Burns, Old JIS 1981, US Lab), Pt 1000 Ohm (DIN/IEC/JIS 1989, Hy-Cal HVAC), Copper 10 and 50 Ω, Ni 120 Ohm (Pure), and Ni 110 (Bristol 7 NA).
• Electrical Connections: Configurable for 2, 3, or 4 wires matching the sensor connections.
• Pulsed Excitation Compatibility: Supports DC to 0.01 second pulse width.

The unit features a high-contrast graphic LCD with LED backlighting and an adjustable digital potentiometer 'DIAL' knob for precise output adjustment in °C, °F, or ohms (resolution of 0.1° or 0.01/0.1 ohm increments). It includes the ability to store up to three preset temperature values.

Power is supplied by four 'AA' Alkaline 1.5V (LR6) batteries, providing approximately 50 hours of battery life with a low-battery indication at roughly one hour remaining. An optional Ni-MH rechargeable kit (Part Number: 020-0103) is available for North America only.

Key Product Differences

The series distinguishes between two primary model configurations:

• CL510A-1: Simulates a single RTD type (Pt100, alpha 0.00385).
• CL510A-10: Supports up to ten different RTD types including various Platinum, Copper, and Nickel standards.

Both models include a Certificate of Calibration traceable to NIST (no points) and carry CE marking for European New Approach Directives. Optional accessories include a rubber boot (Part Number: 020-0209 or CL500A-RB), an RTD Wire Kit (Part Number: 020-0208), and a small carrying case (Part Number: SC-HH500).

• Source One or Multiple RTD Curves 

• High Accuracy 

• Easy-to-Read Backlight Display 

• Calibrate Directly in Temperature (°C or °F) 

Accuracy± 0.015 %
Accuracy Detail± (0.015% of reading + 0.05 Ω)
Battery IndicationA low battery indication is displayed when the battery voltage drops below the operating level
Battery life50 hr
Display TypeLCD
Display Type NoteHigh contrast graphic liquid crystal display, LED backlighting for use in low lit areas
Height45 mm
Length126 mm
Operating Temperature, Max60 °C
Operating Temperature, Min-25 °C
Over-Voltage ProtectionOver-voltage protection to 60 Vdc (rated for 30 sec)
Relative Humidity, Max0.9
Relative Humidity, Min10%
Relative Humidity NoteRelative Humidity Range, 10% ≤RH: ≤90%: 0 to 35°C (32 to 95°F), non-condensing ≤70%: 35 to 60°C (95 to 140°F), non-condensing
Series IDCL510A-Series
Supply Voltage6 Vdc
Temperature Drift± 0.05 Ω/°C
Voltage SourceAccuracy: From 1 to 10.2 mA External Excitation Current: ± (0.015% of rdg + 0.05 Ω) Below 1 mA of External Excitation Current: ± (0.015% of rdg + 0.01 mV/mA excitation current + 0.05 Ω) Resistance Ranges: 0.00 to 410.00, 410.1 to 4001.0 Ω Allowable Excitation Current Range, Steady or Pulsed/Intermittent: <410 Ω: 10.2 mA max 410 to 001 Ω: 1 mA max Pulsed Excitation Current Compatibility: DC to 0.01 second pulse width
Weight0.24 kg
Weight Note0.24 kg (8.4 oz) (including batteries)
Width69 mm