With groundbreaking new proprietary electronics from Fluke Calibration's (patents pending), Metrology Wells let you bring lab-quality performance into whatever field environment you might work in. New analog and digital control techniques provide stability as good as +/- 0.005°C. And with dual-zone control, axial (or “vertical") uniformity is as good as +/- 0.02°C over a 60 mm (2.36 in) zone.

In short, there are six critical components of performance in an industrial heat source: calibrated display accuracy, stability, axial (vertical) uniformity, radial (well-to-well) uniformity, impact from loading, and hysteresis. We added a seventh in the form of a legitimate reference thermometer input and created an entirely new product category: Metrology Wells.

• Best-performing industrial heat sources (accuracy, stability, uniformity) in the world
• Immersion depth to 203 mm (8 in)
• Optional ITS-90 reference input reads PRTs to +/- 0.006 °C
• Four models cover a temperature range from -45°C - 700°C

 

9170

9171

9172

9173

Range
(at 23 °C ambient)

-45°C - 140°C
(-49°F - 284°F)

-30°C - 155°C
(-22°F - 311°F)

35°C - 425°C
(95°F - 797°F)

50°C - 700°C†
(122°F - 1292°F)

Display Accuracy

+/- 0.1°C full range

+/- 0.1°C: 35°C - 100°C
+/- 0.15°C: 100°C - 225°C
+/- 0.2°C: 225°C - 425°C
7nbsp;

+/- 0.2°C: 50°C - 425°C
+/- 0.25°C: 425°C - 660°C

Stability

+/- 0.005°C full range

+/- 0.005°C: 35°C - 100°C
+/- 0.008°C: 100°C - 225°C
+/- 0.01°C: 225°C - 425°C

+/- 0.005°C: 50°C - 100°C
+/- 0.01°C: 100°C - 425°C
+/- 0.03°C: 425°C - 700°C

Axial Uniformity
(60 mm)

+/- 0.1°C at -45°C
+/- 0.04°C at -35°C
+/- 0.02°C at 0°C
+/- 0.07°C at 140°C

+/- 0.025°C at -30°C
+/- 0.02°C at 0°C

+/- 0.07°C at 155°C

+/- 0.05°C: 35°C - 100°C
+/- 0.1°C: 100°C - 225°C
+/- 0.2°C: 225°C - 425°C

+/- 0.1°C: 50°C - 100°C
+/- 0.25°C: 100°C - 425°C

+/- 0.4°C: 425°C - 700°C

Radial Uniformity