5623B Precision Resistance Temperature Detector Freezer Probe

Specially sealed from the sensing element to the end of the probe cable, preventing ingress of moisture when exposed to low temperatures

Have questions? We're here to help!
Have questions? We're here to help!

The 5623B, precision “freezer probe," is specially sealed from the sensing element to the end of the probe cable, preventing ingress of moisture when exposed to temperatures as low as -100°C. The entire assembly withstands temperatures over its full range (-100°C - 156°C), which is ideal for verification of freezers or autoclaves where a thermo-well isn’t available. The 5623B assembly can be fully immersed in fluids when the application may require use in a liquid bath. The 5623B is available in a 6.35 mm (0.25 in) dia. × 125 mm (6 in) long Inconel™ sheath. With accuracy (that includes calibration uncertainty and short-term drift) of +/- 0.05°C over its full range, the 5623B is just right as a secondary standard for calibration of other process sensors.

Most Fluke Calibration readouts make an excellent companion for the 5623B. We recommend the use of the 1523, 1524, 1502A, 1529, or 1560 thermometer readouts.

With each 5623B, you receive a full NVLAP-accredited calibration report. This report includes test data and ITS-90 calibration coefficients to enter into your Fluke Calibration thermometer readout.

• Fully immersible probe assembly to -100°C
• NVLAP-accredited calibration and ITS-90 coefficients included
• Accuracy to +/- 0.05°C over the full range

ResistanceNominal 100 Ω (+/- 0.1 Ω)
Temperature Coefficient0.003925 Ω/Ω/°C nominal
Temperature Range-100°C - 156°C
Transition Temperature-100°C - 156°C
Drift Rate+/- 0.01°C per year maximum at 0°C, when used periodically at max temperature
Calibrated Accuracy (k = 2) +/- 0.05°C over the full range
Cable Length6.1 m (20 ft)
Size6.35 mm dia. x 152 mm (0.25 in dia x 6 in)
Sheath MaterialInconel™ 600
LeadsPTFE-insulated, silver-plated stranded copper, 22 AWG.
TerminationSpecify. See Models and Accessories.
CalibrationIncludes manufacturer's NVLAP-accredited, lab code 200706-0, calibration and table with R vs. T values in 1°C increments from -80°C - 156°C. ITS-90 coefficients included.
  • Thumbnail of document 5623 B User Manual 2331
    Download

    5623 B User Manual 2331

    Download

Request a Service Quote

Get a customized quote for calibration, validation, repair, or training services designed to keep your instruments accurate and compliant.
Learn more

Browse By Brands

Uncover the brands that set the standard for quality and dependability in every industry we serve.
Learn more