The 5616-12 is a 100 Ω platinum resistance thermometer (PRT) with excellent short-term repeatability and comes with an unaccredited NIST-traceable calibration. You won’t find another NIST-traceable reference temperature sensor that matches the accuracy and temperature range of the 5616 for the same price.

The temperature range of the 5616 covers –200°C - 420°C, and its high-purity platinum element and durability make it great for calibrating in the lab or in the field. When choosing a reference with a platinum element, there are two things you want to look at carefully: the short-term repeatability and the long-term drift. When PRTs are thermally cycled over their temperature range as they would be during a calibration, their resistance at the triple point of water can move up and down within an expected range. Fluke Calibration defines this range (called “short-term repeatability) as the repeatability at the triple point of water during three thermal cycles. 5616s are among the best performing in their class with short-term repeatability better than +/- 0.010°C (+/- 0.004°C is typical). In addition, the 5616’s drift is +/- 0.007°C at the triple point of water when exposed up to its maximum temperature (420°C) for 100 hours. These specifications are given at k = 2 and therefore include a 95 % confidence level.

The 5616’s sealed INCONEL® 600 sheath is 298 mm (11.75 in) long and 6.35 mm (0.250 in) in diameter. The probe’s PTFE-jacketed cable is made of silver plated copper that ends with four-wire leads, which eliminate the effects of lead-wire resistance on measurements. Use the 5616 with Fluke Calibration’s 1523/1524 Handheld Reference Thermometer, 1560 Black Stack, 1529 Chub-E4, or 1502A Tweener thermometer readouts.

Each sensor comes with a manufacturer’s report of calibration. The report includes the expanded uncertainty (k = 2) at seven calibration temperature points, ITS-90 calibration coefficients, and a temperature vs. resistance table presented in 1°C increments. Compare the 5616 to other Secondary Reference PRTs. You’ll like its price, but you’ll love its performance.

• Temperature range: –200°C - 420°C
• Excellent stability: +/- 10 mK
• Calibrated accuracy +/- 0.011°C at 0°C
• Reference-grade platinum sensing element
• NIST-traceable calibration included

Temperature Range−200°C - 420°C
Nominal Resistance at 0.01°C100 Ω +/- 0.5 Ω
Temperature Coefficient0.003925 Ω/Ω/°C nominal
Calibrated Accuracy[1] (k = 2)+/- 0.012°C at -200°C
+/- 0.011°C at 0°C
+/- 0.028°C at 420°C
Short-Term Repeatability[2]+/- 0.007°C at 0.010°C
Drift[3]+/- 0.007°C at 0.010°C
Hysteresis+/- 0.010°C maximum
Sensor Length50.8 mm (2.0 in)
Sensor Location9.5 mm +/- 3.2 mm from tip (0.375 in +/- 0.125 in)
Sheath Diameter Tolerance+/- 0.08 mm (+/- 0.003 in)
Sheath MaterialIncontel® 600
Minimum Insulation Resistance500 MΩ at 23°C
Transition Junction Temperature Range[4]−50°C to 150°C (see footnote)
Minimum Immersion Length[5] (< 5 mK error)102 mm (4.0 in)
Maximum Immersion Length254 mm (10 in)
Response Time[5]8 s (typical)
Self Heating (in 0°C Bath)60 mΩ/°C
Lead-Wire Cable TypePTFE-jacketed cable, PTFE insulated conductors, 24 AWG stranded, silver plated copper
Lead-Wire Length182.9 cm +/- 2.5 cm (72.0 in +/- 1.0 in)
Lead-Wire Temperature Range-50°C - 150°C
CalibrationNIST-traceable calibration
[1]Includes calibration uncertainty and 100 hr drift.
[2]Three thermal cycles from min to max temp, includes hysteresis, 95 % confidence (k=2)
[3]After 100 hrs at max temp, 95 % confidence (k=2)
[4]Temperatures outside this range will cause irreparable damage. For best performance, transition junction should not be too hot to touch.
[5]Per ASTM E 644