4180 / 4181 Precision Infrared Calibrators

Fast, accurate and easy-to-use infrared calibrator provides consistent and reliable calibrations

Have questions? We're here to help!
Have questions? We're here to help!

The 4180 Series of Precision Infrared Calibrators for infrared thermometers and thermal imagers is fast, accurate, and easy-to-use. It comes with an accredited calibration from one of the world’s most trusted temperature calibration laboratories, sample calibration procedures for Fluke thermometers built right in and everything you need to get started making high-quality infrared thermometer calibrations. This is the perfect solution for any infrared thermometer or thermal imager within its temperature range.

 

In addition, with accuracies as good as +/-0.35°C the 4180 Series can meet its specifications without additional emissivity-related corrections, leading to legitimate test uncertainty ratios (TUR) as good as 4:1.

• Calibrated radiometrically for meaningful, consistent results.
• Accredited calibration included
• Accurate, reliable performance from –15°C - 500°C
• Large target size of 152 mm (6 in) is required for calibrating most thermometers.
• Light instrument weight of 8.6 kg (19 lbs) makes it easy to lift and carry.
• Intuitive, easy-to-read display that indicates when temperature is stable.

Temperature Range (@ 23 °C Ambient, 0.95 Emissivity)4180: -15°C - 120°C
4181: 35°C - 500°C
Display Accuracy14180:
+/- 0.40°C at -15°C
+/- 0.40°C at 0°C
+/- 0.50°C at 50°C
+/- 0.50°C at 100°C
+/- 0.55°C at 120°C

4181:
+/- 0.35°C at 35°C
+/- 0.50°C at 100°C
+/- 0.70°C at 200°C
+/- 1.20°C at 350°C
+/- 1.60°C at 500°C
Stability4180:
+/- 0.10°C at -15°C
+/- 0.05°C at 0°C
+/- 0.10°C at 120°C

4181:
+/- 0.05°C at 35°C
+/- 0.20°C at 200°C
+/- 0.40°C at 500°C
Uniformity (5.0 in dia of Center of Target)24180:
+/- 0.15°C at -15°C
+/- 0.10°C at 0°C
+/- 0.25°C at 120°C

4181:
+/- 0.10°C at 35°C
+/- 0.50°C at 200°C
+/- 1.00°C at 500°C
Uniformity (2.0 in dia of Center of Target)24180:
+/- 0.10°C at -15°C
+/- 0.10°C at 0°C
+/- 0.20°C at 120°C

4181:
+/- 0.10°C at 35°C
+/- 0.25°C at 200°C
+/- 0.50°C at 500°C
Heating Time4180:
15 min: -15°C - 120°C
14 min: 23°C - 120°C

4181:
20 min: 35°C - 500°C
Cooling Time4180:
15 min: 120°C to 23 °C
20 min: 23°C to –15 °C

4181:
100 min: 500°C to 35°C
40 min: 500°C to 100°C
Stabilization Time10 min
Nominal Emissivity30.95
Thermometer Emissivity Compensation0.9 - 1.0
Target Diameter152.4 mm (6 in)
Computer InterfaceRS-232
Power4180: 115 VAC (+/- 10%), 6.3 A, 50/60 Hz, 630 W
4181: 115 VAC (+/- 10%), 10 A, 50/60 Hz, 1000 W
Fuse(s)4180: 115 VAC, 6.3 A, 250 V, slow blow
4181: 115 VAC, 10 A, 250 V, fast blow
Size (H x W x D)4180: 356 mm x 241 mm x 216 mm (14 in x 9.5 in x 8.5 in)
4181: 356 mm x 241 mm x 216 mm (14 in x 9.5 in x 8.5 in)
Weight4180: 9.1 kg (20 lb)
4181: 9.5 kg (21 lb)
SafetyEN 61010-1:2001, CAN/CSA C22.2 No. 61010.1-04
[1]For 8 µm to 14 µm spectral band thermometers with emissivity set between 0.9 and 1.0
[2]The uniformity specification refers to how IR thermometers with different spot sizes both focused at the center of the targer will measure the same temperature.
[3]The target has a nominal emissivity of 0.95, however it is radiometrically calibrated to minimize emissivity related uncertainties.
  • Thumbnail of document 4180 81omeng0000 14582
    Download

    4180 81omeng0000 14582

    Download
  • Thumbnail of document Cal 210428 3100137 E en 4180 4181 br w 14583
    Download

    Cal 210428 3100137 E en 4180 4181 br w 14583

    Download