Measures and logs thermocouples, PRTs, thermistors, DC voltage, DC current and resistance

Have questions? We're here to help!
Have questions? We're here to help!

The 1586A Super‑DAQ is the most accurate and flexible temperature data acquisition system on the market. It scans and records temperature, DCvoltage, DC current, and resistance for up to 40 input channels and scan speeds as fast as 10 channels per second. The Super-DAQ can be configured for use as a multi-channel data logger in the factory or as a precision reference thermometer for benchtop sensor calibration in the lab.

The 1586A Super-DAQ collects time-stamped precision temperature and electrical measurements for data analysis by technicians, engineers, and quality control personnel to verify process control, analyze interactive systems, ensure conformance to quality standards, or to correlate related events for R&D or troubleshooting. Measurement data and statistics can be viewed in tabular format for all active channels. With the graphing feature, up to four channels can be plotted at the same time, making it easy to quickly assess test setup and results before analyzing the data on a PC. 

When configured with the DAQ-STAQ Multiplexer, the Super-DAQ has the accuracy of the best benchtop reference thermometer readouts for calibration of PRTs, RTDs, thermistors, or thermocouples. Lab efficiency can be increased when the Super-DAQ is connected to a Fluke Calibration drywell or bath and Automated Sensor Test routines are run.

The 1586 is ideal for a number of applications such as thermal mapping, temperature validation, process sensor calibration, and more. These applications are found in a various industries including pharmaceutical, biotechnology, food processing, aerospace, and automotive.

Some temperature data acquisition applications require compliance to government and standards agency requirements. TQSoft and TQAero thermal validation software help engineers, technicians and lab managers to meet the key guidelines for temperature data collection. These software solutions provide a complete qualification, reporting, documentation, and audit trail management system.

• Measure thermocouples, PRTs, thermistors, DC voltage, DC current, and resistance
• Best-in-class temperature measurement accuracy
• Input channels: Up to 40 isolated universal inputs
• Flexible configuration: Internal High-Capacity Module and/or DAQ-STAQ Multiplexer
• Selectable scan speed: Up to 10 channels per second
• Four modes of operation: Scan, Monitor, Measure, Digital Multimeter (DMM)
• Real-time color trending: Chart up to four channels simultaneously
• 6 1/2 digit display resolution for dc voltage, dc current, and resistance
• Current reversal minimizes thermoelectric effects, improving PRT/RTD measurement accuracy
• Automated sensor calibration: Control Fluke Calibration temperature sources such as dry-wells or micro-baths for automated calibration routines
• Data storage: Records up to 20 MB of data and setup files to internal non-volatile memory or to an external USB drive. Transfer data to a PC using the USB drive or LAN connection and view data in Microsoft® Excel
• Data security: Administrator and user profiles for protecting settings and ensuring test traceability
• Mx + B scaling and channel offset zero function
• Alarms: Two independent, user-defined alarms for each channel indicate when an upper or lower range has been exceeded
• Supported by TQSoft and TQAero thermal validation software for FDA 21 CFR Part 11 and AMS 2750 compliant data collection

Mains Voltage100 V Setting: 90 V - 110 V
120 V Setting: 108 V - 132 V
220 V Setting: 198 V - 242 V
240 V Setting: 216 V - 264 V
Frequency47 Hz - 440 Hz
Power Consumption36 VA peak (24 W average)
Environment TemperatureOperating: 0°C - 50°C
Full accuracy: 18°C - 28°C
Storage: −20°C - 70°C
Warm-up: 1 h to full accuracy specifications
Relative Humidity (Non-Condensing)Operating: 0°C - 30°C, <80 %; 30°C - 50°C, <50 %
Storage: −20°C - 70°C, <95 %
AltitudeOperating: 2,000 m
Storage: 12,000 m
Vibration and ShockComplies with MIL-PRF-28800F Class 3
Channel CapacityTotal analog channels: 45
Voltage/resistance channels: 41
Current channels: 5
Digital I/O: 8 bits
Totalizer: 1
Alarm outputs: 6
Trigger input: 1
Channel Input Connector Types1586A Channel 001 (front panel): gold pins, bare wire, spade lugs, banana plugs
DAQ-STAQ Multiplexer: gold pins, bare wire, mini spade lugs, mini banana plugs, mini-jack thermocouple
High-Capacity Module: gold pins, bare wire
Input Protection50 V all functions, terminals and ranges
Math Channels Number of channels: 20
Operations: sum, difference, multiply, divide, polynomial, power, square root, reciprocal, exponential, logarithm, absolute value, average, maximum, minimum
TriggersInterval, external (trigger input), alarm, remote (bus), manual, automated test
MemoryScan data RAM: 75,000 readings with timestamp
Data/Setup flash memory: 20 MB
USB Host PortConnector type: Type A
Function: Memory
File system: FAT32
Memory capacity: 32 GB
USB Device PortConnector type: Type B
Class: Instrument
Function: Control and data transfer
Command protocol: SCPI
LANFunction: Control and data transfer
Network protocols: Ethernet 10/100, TCP/IP
Command protocol: SCPI
RS-232 Connector: D-sub 9 pin (DE-9)
Baud rates: 1200, 2400, 4800, 9600, 19200, 38400
Function: Temperature source control output
Dimensions (H x W x D)150 mm x 245 mm x 385 mm
Weight: 6 kg (typical configuration)
Shipping Weight9.5 kg (typical configuration)
ConformityCE, CSA, IEC 61010 3rd ed.

General Specifications

Mains Voltage

100 V Setting: 90 V to 110 V
120 V Setting: 108 V to 132 V
220 V Setting: 198 V to 242 V
240 V Setting: 216 V to 264 V

Frequency

47 Hz to 440 Hz

Power Consumption 

36 VA peak (24 W average)

Environment Temperature

Operating: 0 °C to 50 °C
Full accuracy: 18 °C to 28 °C
Storage: −20 °C to 70 °C
Warm-up: 1 hour to full accuracy specifications

Relative Humidity (non-condensing)

Operating:  0 °C to 30 °C <80 %;
                   30 °C to 50 °C <50 %
Storage: −20 °C to 70 °C <95 %

Altitude

Operating: 2,000 m
Storage: 12,000 m

Vibration and Shock 

Complies with MIL-PRF-28800F Class 3

Channel Capacity

Total analog channels: 45
Voltage/resistance channels: 41
Current channels:  5
Digital I/O: 8 bits
Totalizer: 1
Alarm outputs: 6
Trigger input: 1

Input Protection 

50 V all functions, terminals and ranges

Math Channels

Number of channels: 20
Operations: sum, difference, multiply, divide, polynomial, power, square root, reciprocal, exponential, logarithm, absolute value, average, maximum, minimum

Triggers

interval, external (trigger input), alarm, remote (bus), manual, automated test

Memory

Scan data RAM: 75,000 readings with timestamp
Data/Setup flash memory: 20 MB

USB Host Port

Connector type: Type A
Function: Memory
File system: FAT32
Memory capacity: 32 GB

USB Device Port

Connector type: Type B
Class: Instrument
Function: Control and data transfer
Command protocol: SCPI

LAN

Function: Control and data transfer
Network protocols: Ethernet 10/100, TCP/IP
Command protocol: SCPI

RS-232

Connector: D-sub 9 pin (DE-9)
Baud rates: 1200, 2400, 4800, 9600, 19200, 38400
Function: Temperature source control output

  • Thumbnail of document Fluke 1586 Data Sheet 2315
    Download

    Fluke 1586 Data Sheet 2315

    Download
  • Thumbnail of document 1586 A cmeng0000 14587
    Download

    1586 A cmeng0000 14587

    Download

Accessories

Request a Service Quote

Get a customized quote for calibration, validation, repair, or training services designed to keep your instruments accurate and compliant.
Learn more

Browse By Brands

Uncover the brands that set the standard for quality and dependability in every industry we serve.
Learn more