You need accuracy for compliance, product yields, energy savings, and consistent results. The Fluke 1523 and Fluke 1524 use current reversal, a technique used in high-end instruments that eliminates thermal EMFs, for precision temperature measurements. Specifications are guaranteed from -10ºC - 60ºC ambient. Special precision resistors and a highly stable reference voltage source keep 1523 and 1524 accuracy virtually insensitive to environmental temperature.

The Fluke 1523 and Fluke 1524 Reference Thermometers from Fluke Calibration measure, graph, and record PRTs, thermocouples, and thermistors. These thermometer readouts deliver exceptional accuracy, wide measurement range, temperature logging, and trending, all in a handheld tool you can take anywhere.

The Fluke 1523 and Fluke 1524 let you handle field applications, laboratory measurements, and data logging with ease. And with the dual channel measurement capabilities of the model Fluke 1524, you can do twice the work in half the time.

Like all Fluke handheld tools, the 1523 and 1524 Reference Thermometers endure rigorous testing in temperature extremes and under harsh conditions of vibration, so you can take them anywhere you need to go.

 

• Three sensor types: PRTs, Thermocouples, and Precision Thermistors
• High Accuracy: PRTs (up to +/- 0.011°C), Thermocouples (+/- 0.24°C for J,K,L,M), Precision Thermistors (+/- 0.002°C)
• Fast mode: PRTs (0.45 s per sample), Thermocouples (0.3 s per sample), Precision Thermistors (0.3 s per sample)
• Graphical display with a 128 x 64 backlit LCD graphic display
• Plot and scale trends in real time
• Simultaneous dual channel readings

Input Channels1523: 1
1524: 2
Logging1523: 25 readings with statistics
1524: 15,000 time and date stamped; 25 readings with statistics
Sample Interval (Normal)1 s
Sample Interval (Fast Mode)0.3 s (see technical manual for details)
Sensor TypesPRTs, RTDs, Thermistors, and Thermocouples
Thermocouple TypesC, E, J, K, L, M, N, T, U, B, R, S
Operating Temperature-10ºC - 60ºC (best accuracy 13ºC - 33ºC)
Power Requirements3 AA alkaline batteries
Size96 mm x 200 mm x 47 mm (3.75 in x 7.9 in x 1.86 inches)
Weight0.65 kg (1.4 lb)
Environmental Conditions for Best Accuracy13°C - 33°C
Millivolt Range and Accuracy-10 mV - 75 mV +/- (0.005% + 5 μV)
Resistance Range and Accuracy0 Ω - 400 Ω +/- (0.004% + 0.002 Ω)
200 Ω - 50 kΩ +/- (0.01% + 0.5 Ω)
50 kΩ - 500 kΩ +/- (0.03%)
Temperature Coefficient, Voltage (-10°C - 13°C , 33°C - 60°C)+/- (0.001%/°C + 1 μV/°C)
Temperature Coefficient, Resistance (-10°C - 13°C , 33°C - 60°C)0.0008%/°C + 0.0004 Ω (0 Ω - 400 Ω)
0.002%/°C + 0.1 Ω (0 Ω - 50 kΩ)
0.06 %/°C + 0.1 Ω(50 kΩ to 500 kΩ)
Excitation Current, Resistance1 mA (0 Ω - 400 Ω)
10 μA (0 Ω - 50 kΩ)
2 µA (50 kΩ to 500 kΩ)
  • Thumbnail of document 3383390c en 1523 1524 reference thermometers w 14584
    Download

    3383390c en 1523 1524 reference thermometers w 14584

    Download
  • Thumbnail of document 1523 24 tgeng0000 14585
    Download

    1523 24 tgeng0000 14585

    Download

Request a Service Quote

Get a customized quote for calibration, validation, repair, or training services designed to keep your instruments accurate and compliant.
Learn more

Browse By Brands

Uncover the brands that set the standard for quality and dependability in every industry we serve.
Learn more