1200 Two Pressure Humidity Generator

Produces accurate humidity values using the fundamental, NIST proven, two-pressure principle, 0.5 %RH uncertainty

This product is no longer available. Contact us at 800-567-8686 for alternative solutions.

Have questions? We're here to help!
Have questions? We're here to help!

The Model 1200 Mini Humidity Generator produces accurate humidity values using the fundamental, NIST proven, “two-pressure” principle. The 1200 will automatically supply relative humidity, dew point, frost point, and other calculated values for instrument calibration and evaluation as well as precision environmental testing. This system automatically generates multipoint profiles as well as manually entered humidity levels, while continuously storing and printing system data.

Virtually all functions of the 1200 humidity generator are computer controlled. All desired humidities, temperatures, and time intervals may be programmed. Visual indications of system status are displayed in real time on the computer screen. The automated features of the 1200 allow the generation of known humidity levels completely unattended. This frees the operating technician from the task of monitoring and adjusting.

 

• 0.5 %RH Uncertainty
• Traceable to SI
• Based on NIST Proven Two-Pressure Principle
• Generate: RH, DP, FP, PPM, Multipoint Profiles
• Computerized Internal Transducer Calibration
• Computes System Uncertainties in Real Time
• Automatically Applies Enhancement Factors
• No Refrigerants - Thermoelectric Cooling/Heating
• Only 4 square feet of floor space (20" x 30 ")
• Touch-screen control
• USB and Ethernet Interface
• Oil-Less Compressed Air System w/Membrane Dryer
• 1200 Utility Cart with Power Strip

Relative Humidity Range10 to 95%
Relative Humidity Uncertainty0.5%
Frost Point Temperature Range–18 to 0 °C
Dew Point Temperature Range–20 to 50 °C
Dew Point Uncertainty0.1 °C
Chamber Temperature Range10 to 60 °C
Chamber Temperature Control Stability±0.04 °C
Chamber Temperature Uniformity0.1 °C
Chamber Temperature Measurement Uncertainty0.05 °C
Chamber Temperature Cooling Rate4 Minutes Per °C Average
Chamber Temperature Heating Rate2 Minutes Per °C Average
Saturation Pressure RangeAmbient to 152 psiA
Saturation Pressure Uncertainty0.08% of FS psiA
Test Chamber Pressure RangeAmbient
Test Chamber Pressure Uncertainty0.08% of FS psiA
Display Resolution0.01
Gas TypeAir or Nitrogen
Gas Pressure Rating (MAWP)175 psiG
Gas Flow Rate Resolution0.01 L/min
Gas Flow Rate Uncertainty1.0 L/min
Test Chamber Dimensions6" x 6" x 6" (152 mm x 152 mm x 152 mm)
Access PortØ1.688" (42.8 mm) located on right side
Physical Dimensions11.85" H x 24.10" W x 14.11" D (30.1 cm x 61.2 cm x 35.8 cm)
Dry Weight (Generator Only)56 lbs. (25.40 Kg)
Wet Weight (Generator Only)65 lbs. (29.48 Kg)
Utilities
Electrical Power100/240 V, ~6/3 A, 50/60 Hz
Gas Supply155-175 psiG @ 0.5 cfm (15 L/min)
Environmental
Operating Temperature15 to 30 °C
Storage Temperature0 to 50 °C
Humidity5 to 95% Non-condensing
Gas Flow Rate Range2 to 10 L/min

Virtually any humidity and temperature may be generated within the operational limits of the generator. The output or recording of the device under test may then be compared with the generator’s data for analysis.

Chilled Mirror Hygrometers: Install the actual chilled mirror head into the chamber or insert a sample tube through the test port and draw a sample through the chilled mirror head and you can: verify mirror temperature measurement accuracy (calibration) when the hygrometer is in thermal equilibrium with its environment; perform operational checks of the heat-pump and optical components before and after mirror cleaning and balancing; determine whether the hygrometer is controlling the mirror deposit in the liquid phase or ice phase when operating at dew and frost points below 0 °C; determine if the hygrometer is correctly calculating other humidity parameters; determine the hygrometer’s repeatability, stability, and drift characteristics.

Humidity Sensors and Data Loggers: Insert your humidity probes through the chamber access port or install the data logger into the chamber and you can: determine humidity calibration accuracy and/or characterize humidity sensitivity by subjecting the humidity sensor to a variety of humidity levels; perform operational checks such as the
sensing systems capability to correctly calculate and display other humidity parameters; determine the repeatability, stability, hysteresis, and drift characteristics of various humidity sensing systems.

Environmental Testing: The 1200 can serve as a test bed for evaluation and R&D of humidity sensors, humidity sensing systems, and humidity sensitive products, e.g., polymers, composites, film, magnetic medium, pharmaceuticals, soil hydrology, consumables, electronics, optics, etc.
 

  • Thumbnail of document 1200 Mini Data Sheet 10176
    Download

    1200 Mini Data Sheet 10176

    Download
  • Thumbnail of document 1200 Mini Manual 10177
    Download

    1200 Mini Manual 10177

    Download

Request a Service Quote

Get a customized quote for calibration, validation, repair, or training services designed to keep your instruments accurate and compliant.
Learn more

Browse By Brands

Uncover the brands that set the standard for quality and dependability in every industry we serve.
Learn more